Metformin on Force FluoroPhenyl by UHPLC-MS/MS

LC_CF0671
PeakstR (min)Precursor IonProduct Ion Product Ion
1.Metformin1.00130.371.060.1
ColumnForce FluoroPhenyl (cat.# 9639252)
Dimensions:50 mm x 2.1 mm ID
Particle Size:1.8 µm
Temp.:40 °C
Sample
Diluent:Acetonitrile
Conc.:100 ng/mL
Inj. Vol.:2 µL
Mobile Phase
A:0.1% Formic acid in water
B:0.1% Formic acid in acetonitrile
Time (min)Flow (mL/min)%A%B
0.000.63070
1.50.69010
1.510.63070
2.50.63070
DetectorMS/MS
Ion Mode:ESI+
Mode:MRM
InstrumentUHPLC