Welcome to the New Restek.com! Take a tour of the new site and feel free to send us your feedback.
Your web browser will no longer be supported by Restek.com as of 30 June 2021.
To avoid any interruption in access or functionality, install a current-generation web browser now. Learn more.

Cardiac Drugs on Force Biphenyl by LC-MS/MS

LC_CF0675
PeakstR (min)Precursor IonProduct IonProduct IonPolarity
1.Digoxin0.95798.4651.4355.1+
2.Flecainide1.28415.3397.9300.9+
3.Furosemide1.32329.0205.2285.1-
4.Verapamil1.68455.1165.2150.0+
5.Digitoxin1.85782.4635.2243.1+
6.Amiodarone3.17646.158.186.2+
ColumnForce Biphenyl (cat.# 9629252)
Dimensions:50 mm x 2.1 mm ID
Particle Size:1.8 µm
Temp.:40 °C
Sample
Diluent:Water:acetonitrile (90:10)
Conc.:100 ng/mL
Inj. Vol.:2 µL
Mobile Phase
A:Water + 5 mM ammonium formate + 0.1% formic acid
B:Acetonitrile + 0.1% formic acid
Time (min)Flow (mL/min)%A%B
0.000.36040
3.500.3595
3.510.36040
5.50.36040
DetectorMS/MS
Ion Mode:ESI+/ESI-
Mode:Scheduled MRM
InstrumentUHPLC