Welcome to the New Restek.com! Take a tour of the new site and feel free to send us your feedback.
Your web browser will no longer be supported by Restek.com as of 30 June 2021.
To avoid any interruption in access or functionality, install a current-generation web browser now. Learn more.

Morphine and Metabolites on Force Biphenyl by UHPLC-MS/MS

LC_CF0677
PeakstR (min)Precursor IonProduct IonProduct Ion
1.Morphine-3-β-D-glucuronide0.70462.1286.0152.1
2.Morphine1.26286.1164.9152.0
3.Morphine-6-β-D-glucuronide1.34462.1286.0152.1
4.Morphine-N-oxide1.41302.1284.9161.9
5.6-Acetylmorphine1.86328.1211.1165.0
ColumnForce Biphenyl (cat.# 9629252)
Dimensions:50 mm x 2.1 mm ID
Particle Size:1.8 µm
Temp.:35 °C
Sample
Diluent:0.1% Formic acid in water
Conc.:50 ng/mL
Inj. Vol.:5 µL
Mobile Phase
A:0.1% Formic acid in water
B:0.1% Formic acid in methanol
Time (min)Flow (mL/min)%A%B
0.000.58515
0.500.58515
2.000.53070
2.010.58515
4.000.58515
DetectorMS/MS
Ion Mode:ESI+
Mode:MRM
InstrumentUHPLC