Warfarin and Chemotherapeutic Drugs on Raptor FluoroPhenyl by LC-MS/MS

LC_CF0678
PeakstR (min)Conc.
(ng/mL)
Precursor IonProduct IonProduct Ion
1.Carboplatin0.83100372.00294.00355.00
2.5-Fluorouracil0.89100130.98113.9957.99
3.Methotrexate2.1820455.15308.09175.14
4.Ifosfamide2.6720261.0391.95153.90
5.Cyclophosphamide2.7520261.03139.97106.01
6.Etoposide3.10100589.34229.01185.03
7.Warfarin3.5810309.11162.99251.01
ColumnRaptor FluoroPhenyl (cat.# 9319A12)
Dimensions:100 mm x 2.1 mm ID
Particle Size:2.7 µm
Pore Size:90 Å
Guard Column:Raptor FluoroPhenyl EXP guard column cartridge 5 mm, 2.1 mm ID, 2.7 µm (cat.# 9319A0252)
Temp.:50 °C
Sample
Diluent:Water:methanol (95:5)
Conc.:10-100 ng/mL
Inj. Vol.:10 µL
Mobile Phase
A:0.1% Formic acid in water
B:Methanol
Time (min)Flow (mL/min)%A%B
0.000.4955
4.000.40100
4.010.4955
6.000.4955
Max Pressure:400 bar
DetectorMS/MS
Ion Mode:ESI+
Mode:MRM
InstrumentUHPLC
NotesThe sample solution was prepared by protein precipitation using Resprep PPT3 96-well plates (cat# 26489) and reservoir (cat# 26493). A 200 μL aliquot of fortified human plasma was mixed with 600 μL of acetonitrile by vortexing for 30 seconds at 2,000 rpm. The sample was filtered using a vacuum manifold. The filtrate was dried and reconstituted with 100 μL of water:methanol (95:5) and injected for analysis.