Welcome to the New Restek.com! Take a tour of the new site and feel free to send us your feedback.
Your web browser will no longer be supported by Restek.com as of 30 June 2021.
To avoid any interruption in access or functionality, install a current-generation web browser now. Learn more.

Bile Acids in Human Plasma on Raptor C18 (1.8 μm) by LC-MS/MS

LC_CF0711
PeakstR (min)Conc.
(ng/mL)
Precursor IonProduct Ion
1.Glycoursodeoxycholic acid (GUDCA)1.451540448.474.1
2.Tauroursodeoxycholic acid (TUDCA)1.5901,080498.480.1
3.Glycocholic acid (GCA)2.012540464.374.2
4.Taurocholic acid (TCA)2.1591,080514.480.0
5.Ursodeoxycholic acid (UDCA)2.254360391.4391.4
6.Hyodeoxycholic acid (HDCA)2.560360391.4391.4
7.Cholic acid (CA)2.74190407.3407.2
8.Glycochenodeoxycholic acid (GCDCA)3.293540448.474.1
PeakstR (min)Conc.
(ng/mL)
Precursor IonProduct Ion
9.Taurochenodeoxycholic acid (TCDCA)3.4171,080498.480.1
10.Glycodeoxycholic acid (GDCA)3.540540448.474.1
11.Taurodeoxycholic acid (TDCA)3.6531,080498.480.1
12.Chenodeoxycholic acid (CDCA)4.197360391.4391.4
13.Deoxycholic acid (DCA)4.39990391.4391.4
14.Glycolithocholic acid (GLCA)4.567540432.374.0
15.Taurolithocholic acid (TLCA)4.6861,080482.480.0
16.Dehydrolithocholic acid (DHLCA)5.61890373.3373.3
17.Lithocholic acid (LCA)5.763180375.5375.3
ColumnRaptor C18 (cat.# 9304252)
Dimensions:50 mm x 2.1 mm ID
Particle Size:1.8 µm
Pore Size:90 Å
Guard Column:UltraShield UHPLC precolumn filter 0.2 µm (cat.# 25810)
Temp.:60 °C
Sample
Diluent:70:30 Water:methanol
Inj. Vol.:3 µL
Mobile Phase
A:5 mM Ammonium acetate in water
B:50:50 Acetonitrile:methanol
Time (min)Flow (mL/min)%A%B
0.000.56535
2.000.56040
2.500.55545
3.500.55050
4.600.54555
5.700.52080
5.900.8*585
6.500.8*585
6.510.56535
8.500.56535
DetectorMS/MS
Ion Mode:ESI-
Mode:MRM
InstrumentUHPLC
Notes*The flow rate was increased to 0.8 mL/min to more thoroughly flush phospholipids from the analytical column, thereby reducing matrix effects.