Welcome to the New Restek.com! Take a tour of the new site and feel free to send us your feedback.
Your web browser will no longer be supported by Restek.com as of 30 June 2021.
To avoid any interruption in access or functionality, install a current-generation web browser now. Learn more.

Psilocybin and Psilocin on Raptor HILIC-Si by LC-MS/MS

LC_CF0764
PeakstR (min)Conc.
(ng/mL)
Precursor IonProduct Ion 1Product Ion 2
1.Psilocin1.12100205.1160.1115.0
2.Psilocybin1.73100285.1240.0205.2
ColumnRaptor HILIC-Si (cat.# 9310A52)
Dimensions:50 mm x 2.1 mm ID
Particle Size:2.7 µm
Pore Size:90 Å
Temp.:40 °C
Sample
Diluent:10 mM Ammonium formate in 90:10 acetonitrile:water (v/v)
Conc.:100 ng/mL
Inj. Vol.:2 µL
Mobile Phase
A:Water, 10 mM ammonium formate
B:90:10 Acetonitrile:water (v/v), 10 mM ammonium formate
Time (min)Flow (mL/min)%A%B
0.000.50100
0.20.50100
1.700.5955
1.710.50100
3.000.50100
DetectorMS/MS
Ion Source:Electrospray
Ion Mode:ESI+
Mode:MRM
NotesA 100 ng/mL standard mix of psilocin and psilocybin was prepared in 10 mM ammonium formate in 90:10 acetonitrile:water (v/v). The solution was vortexed at 3000 rpm for 10 seconds to mix, and then the supernatant was injected for LC-MS/MS analysis.