Your web browser will no longer be supported by Restek.com as of 30 June 2021.
To avoid any interruption in access or functionality, install a current-generation web browser now. Learn more.
Peaks | tR (min) | Conc. (ng/mL) | Precursor | Product 1 | Product 2 | |
---|---|---|---|---|---|---|
1. | Phenobarbital | 2.01 | 500 | 230.9 | 42.1 | - |
2. | Butalbital | 2.14 | 500 | 223.0 | 180.2 | 42.1 |
3. | Amobarbital | 2.56 | 500 | 225.0 | 182.2 | 42.1 |
4. | Pentobarbital | 2.62 | 500 | 225.0 | 182.2 | 42.1 |
5. | Secobarbital | 2.81 | 500 | 337.0 | 194.2 | 42.1 |
6. | THC-COOH | 3.26 | 5 | 343.0 | 299.2 | 245.1 |
7. | THCA-A | 3.58 | 5 | 357.1 | 313.3 | 245.1 |
Column | Force Biphenyl (cat.# 962935E) | ||||||||||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Dimensions: | 50 mm x 3.0 mm ID | ||||||||||||||||||||||||||||
Particle Size: | 3 µm | ||||||||||||||||||||||||||||
Pore Size: | 100 Å | ||||||||||||||||||||||||||||
Guard Column: | Force Biphenyl EXP guard column cartridge 5 mm, 3 mm ID, 5 µm (cat.# 962950253) | ||||||||||||||||||||||||||||
Temp.: | 30 °C | ||||||||||||||||||||||||||||
Standard/Sample | Phenobarbital (cat.# 34037) | ||||||||||||||||||||||||||||
Pentobarbital (cat.# 34036) | |||||||||||||||||||||||||||||
Butalbital (cat.# 34032) | |||||||||||||||||||||||||||||
Secobarbital (cat.# 34038) | |||||||||||||||||||||||||||||
(±)11-Nor-9-carboxy-D9-THC (cat.# 34068) | |||||||||||||||||||||||||||||
d9-Tetrahydrocannabinolic acid A (cat.# 34111) | |||||||||||||||||||||||||||||
Diluent: | 1:9 Urine:water | ||||||||||||||||||||||||||||
Inj. Vol.: | 1 µL | ||||||||||||||||||||||||||||
Mobile Phase | |||||||||||||||||||||||||||||
A: | Water, 0.1% formic acid | ||||||||||||||||||||||||||||
B: | Methanol, 0.1% formic acid | ||||||||||||||||||||||||||||
|
Detector | MS/MS |
---|---|
Ion Source: | Electrospray |
Ion Mode: | ESI- |
Sample Preparation | A 100 μL aliquot of urine was spiked with analytes and diluted with 900 μL of MS water in a 2 mL screw-thread vial (cat.# 21143) and capped with a short-cap, screw-vial closure (cat.#24498). The sample was vortexed for 10 seconds and injected for LC-MS/MS analysis. |
Notes | The analytical and guard columns were equipped with an UltraShield precolumn filter (cat. 25809). |