Welcome to the New Restek.com! Take a tour of the new site and feel free to send us your feedback.

Your web browser will no longer be supported by Restek.com as of 30 June 2021.
To avoid any interruption in access or functionality, install a current-generation web browser now. Learn more.

EPA 8310 PAH Mix on Pinnacle II PAH (FLD)

ColumnPinnacle II PAH, custom (cat.# 921946E)
Dimensions:150 mm x 3.0 mm ID
Particle Size:4 µm
Pore Size:110 Å
Temp.:30 °C
SampleEPA Method 8310 PAH mixture (cat.# 31841)
Diluent:Acetonitrile
Conc.:30 ng/mL (ppb)
Inj. Vol.:5 µL
Mobile Phase
A:Water
B:Acetonitrile
Time (min)Flow (mL/min)%A%B
0.001.15545
6.501.14060
8.001.13070
12.001.10100
15.001.10100
15.011.15545
17.001.15545
DetectorShimadzu RF-10AXL
Time (min)Wavelength Ex (nm)Wavelength Em (nm)Gain
027033016x
7.5025037016x
9.203304304x
11.002703901x
12.202904301x
14.703704604x
InstrumentShimadzu Prominence
NotesRF-10AXL sensitivity: medium
Acenaphthylene is included in PAH standard mix (cat.# 31841) but does not fluoresce.