EPA 8310 PAH Mix on Pinnacle II PAH (FLD)

ColumnPinnacle II PAH, custom (cat.# 921946E)
Dimensions:150 mm x 3.0 mm ID
Particle Size:4 µm
Pore Size:110 Å
Temp.:30 °C
SampleEPA Method 8310 PAH mixture (cat.# 31841)
Diluent:Acetonitrile
Conc.:30 ng/mL (ppb)
Inj. Vol.:5 µL
Mobile Phase
A:Water
B:Acetonitrile
Time (min)Flow (mL/min)%A%B
0.001.15545
6.501.14060
8.001.13070
12.001.10100
15.001.10100
15.011.15545
17.001.15545
DetectorShimadzu RF-10AXL
Time (min)Wavelength Ex (nm)Wavelength Em (nm)Gain
027033016x
7.5025037016x
9.203304304x
11.002703901x
12.202904301x
14.703704604x
InstrumentShimadzu Prominence
NotesRF-10AXL sensitivity: medium
Acenaphthylene is included in PAH standard mix (cat.# 31841) but does not fluoresce.