Explosives on Raptor Biphenyl by EPA Method 8330B

ColumnRaptor Biphenyl (cat.# 9309A65)
Dimensions:150 mm x 4.6 mm ID
Particle Size:2.7 µm
Temp.:35 °C
Sample8330B Nitroaromatics and nitramine mix (cat.# 33204)
1,4-Dinitrobenzene (cat.# 33205)
Diluent:Water
Conc.:20 µg/mL
Inj. Vol.:5 µL
Mobile Phase
A:Water
B:Methanol
Time (min)Flow (mL/min)%A%B
0.001.03070
101.03070
DetectorUV/Vis @ 210/254 nm
Cell Temp:35 °C
Flow Cell Size:Semi-micro
InstrumentShimadzu Prominence
NotesThis is the recommended primary column for a two-column analysis. See LC_EV0530 for the confirmation column.