Welcome to the New Restek.com! Take a tour of the new site and feel free to send us your feedback.
Your web browser will no longer be supported by Restek.com as of 30 June 2021.
To avoid any interruption in access or functionality, install a current-generation web browser now. Learn more.

Ultrashort- Through Long-Chain and Alternative PFAS on Raptor Polar X

LC_EV0569
PeakstR (min)Conc.
(ng/L)
Precursor IonProduct Ion
1.11-Chloroeicosafluoro-3-oxanonane-1-sulfonate (11CL-PF3OUdS)1.25400630.78450.80
2.9-Chlorohexadecafluoro-3-oxanonane-1-sulfonate (9Cl-PF3ONS)1.34400530.78350.85
3.Perfluorooctanesulfonic acid (PFOS)1.38400498.8479.97
4.Perfluorohexanesulfonic acid (PFHxS)1.49400398.9079.97
5.Perfluorobutanesulfonic acid (PFBS)1.64400298.9779.97
6.Perfluoropropanesulfonic acid (PFPrS)1.73400248.9779.98
7.Perfluoroethanesulfonic acid (PFEtS)1.86400198.9879.92
PeakstR (min)Conc.
(ng/L)
Precursor IonProduct Ion
8.Hexafluoropropylene oxide dimer acid (HFPO-DA)2.06400284.97168.92
9.Perfluorooctanoic acid (PFOA)2.11400412.90368.91
10.Ammonium 4,8-dioxa-3H-perfluorononanoate (ADONA)2.15400376.90250.93
11.Perfluorohexanoic acid (PFHxA)2.36400312.97268.90
12.Perfluorobutanoic acid (PFBA)2.76400212.97168.97
13.Perfluoropropionic acid (PFPrA)3.06400163.03119.01
14.Trifluoroacetic acid (TFA)3.77400113.0369.01
ColumnRaptor Polar X (cat.# 9311A52)
Dimensions:50 mm x 2.1 mm ID
Particle Size:2.7 µm
Temp.:40 °C
Sample
Diluent:50:50 Water:methanol
Conc.:400 ng/L
Inj. Vol.:10 µL
Mobile Phase
A:Water, 10 mM ammonium formate, 0.05% formic acid
B:60:40 Acetonitrile:methanol, 0.05% formic acid
Time (min)Flow (mL/min)%A%B
0.000.51585
8.000.51585
DetectorMS/MS
Ion Mode:ESI-
Mode:MRM
InstrumentUHPLC
Notes