Your web browser will no longer be supported by Restek.com as of 30 June 2021.
To avoid any interruption in access or functionality, install a current-generation web browser now. Learn more.

EPA Method 533 PFAS Calibration Standard

LC_EV0584
PeakstR (min)Conc.
(ng/mL)
Precursor IonProduct Ion
1.Perfluoro-n-butanoic acid (PFBA)1.93325213169
2.Perfluoro-3-methoxypropanoic acid (PFMPA)2.5442522985
3.Perfluoro-n-pentanoic acid (PFPeA)3.36925263219
4.Perfluorobutanesulfonic acid (PFBS)3.5942529980
5.Perfluoro-4-methoxybutanoic acid (PFMBA)3.6692527985
6.Perfluoro(2-ethoxyethane)sulfonic acid (PFEESA)3.91925315135
7.Perfluoro-3,6-dioxaheptanoic acid (NFDHA)4.08425295201
8.1H,1H,2H,2H-Perfluorohexane sulfonic acid (4:2 FTS)4.12925327307
9.Perfluorohexanoic acid (PFHxA)4.18925313269
10.Perfluoro-1-pentanesulfonic acid (PFPeS)4.2782534980
11.Perfluoro(2-methyl-3-oxahexanoic) acid (HFPO-DA)4.36525285169
12.Perfluoroheptanoic acid (PFHpA)4.71525363319
PeakstR (min)Conc.
(ng/mL)
Precursor IonProduct Ion
13.Perfluoro-1-hexanesulfonic acid (PFHxS)4.7502539980
14.4,8-dioxa-3H-perfluorononanoic acid (ADONA)4.7825277251
15.1H,1H,2H,2H-Perfluorooctane sulfonic acid (6:2 FTS)5.09625427407
16.Perfluoro-1-heptanesulfonic acid (PFHpS)5.1322544980
17.Perfluorooctanoic acid (PFOA)5.11525413369
18.Perfluorooctanesulfonic acid (PFOS)5.4412549980
19.Perfluorononanoic acid (PFNA)5.43925463419
20.9-Chlorohexadecafluoro-3-oxanonane-1-sulfonic acid (9Cl-PF3ONS)5.58825531351
21.1H,1H,2H,2H-Perfluorodecane sulfonic acid (8:2 FTS)5.71225527507
22.Perfluorodecanoic acid (PFDA)5.71225513469
23.Perfluoroundecanoic acid (PFUnA)5.95425563519
24.11-Chloroeicosafluoro-3-oxaundecane-1-sulfonic acid (11Cl-PF3OUdS)6.04925631451
25.Perfluorododecanoic acid (PFDoA)6.15825613569
ColumnForce C18 (cat.# 9634252)
Dimensions:50 mm x 2.1 mm ID
Particle Size:1.8 µm
Pore Size:100 Å
Temp.:40 °C
Standard/SampleEPA 533 PFAS calibration standard (cat.# 30736)
Diluent:80:20 Methanol:water
Conc.:25 ng/mL
Inj. Vol.:3 µL
Mobile Phase
A:Water, 5 mM ammonium acetate
B:Methanol
Time (min)Flow (mL/min)%A%B
0.000.47030
6.000.4595
6.600.4595
6.610.48020
7.500.48020
DetectorShimadzu 8045
Ion Source:Electrospray
Ion Mode:ESI-
Mode:MRM
InstrumentShimadzu Nexera X2
NotesBranched isomers for PFOA, PFOS, and PFHxS labeled as peak number "a" and "b".
PFAS delay column used (cat.# 27854).