Welcome to the New Restek.com! Take a tour of the new site and feel free to send us your feedback.
Your web browser will no longer be supported by Restek.com as of 30 June 2021.
To avoid any interruption in access or functionality, install a current-generation web browser now. Learn more.

Pesticides on Pinnacle DB Aqueous C18

LC_FF0479B
PeakstR (min)PolarityMol. Wt. Transition 1
1.Bentazone0.77negative240.28239.0→131.8
2.Bromoxynil1.32negative276.91275.9→80.8
3.Ioxynil2.46negative370.92369.8→126.7
4.Fludioxonil5.34negative248.19247.0→179.9
5.Boscalid5.36negative343.21343.0→307.0
6.Prothioconazole5.41negative344.26341.9→305.8
7.Fipronil5.56negative437.15434.9→329.8
8.Noviflumuron7.08negative529.1527.1→343.8
ColumnPinnacle DB Aqueous C18 (cat.# 9418252)
Dimensions:50 mm x 2.1 mm ID
Particle Size:1.9 µm
Pore Size:140 Å
Temp.:35 °C
Samplemulticomponent pesticide standard
Diluent:water
Conc.:33.3 ppb each pesticide
Inj. Vol.:5 µL
Mobile Phase
A:10 mM NH4OAc in water
B:10 mM NH4OAc in methanol
Time (min)%B
0.0010
110
890
1090
1110
Flow:0.60 mL/min
Max Pressure:~517 bar
DetectorApplied Biosystems/MDS Sciex LC/MS/MS
Model #:4000 QTRAP LC-MS/MS system
Ion Source:TurboIonSpray®
Ion Mode:ESI-
Ion Spray Voltage:-4.2 kV
Gas 1:40 psi (275.8 kPa)
Gas 2:60 psi (413.7 kPa)
Source Temp.:500 °C
InstrumentShimadzu UFLCXR