Welcome to the New Restek.com! Take a tour of the new site and feel free to send us your feedback.
Your web browser will no longer be supported by Restek.com as of 30 June 2021.
To avoid any interruption in access or functionality, install a current-generation web browser now. Learn more.

Pesticides on Ultra Aqueous C18

LC_FF0490B
PeakstR (min)Transition 1Transition 2
1.Methamidophos1.06142 -> 94142 -> 125
2. Acephate1.27184.1 -> 143184.1 -> 125
3.Omethoate1.69214 -> 124.9214 -> 182.8
4.Propamocarb6.61189.2 -> 102.2189.2 -> 73.9
ColumnUltra Aqueous C18 (cat.# 9178312)
Dimensions:100 mm x 2.1 mm ID
Particle Size:3 µm
Pore Size:100 Å
Temp.:35 °C
Samplemulticomponent pesticide standard
Diluent:water
Conc.:1 ppb each pesticide
Inj. Vol.:10 µL
Mobile Phase
A:10 mM NH4OAc in water
B:10 mM NH4OAc in methanol
Time (min)%B
0.0020
8.090
12.0100
14.8100
14.920
Flow:0.5 mL/min
DetectorApplied Biosystems/MDS Sciex LC/MS/MS
Model #:4000 QTRAP
Ion Source:TurboIonSpray®
Ion Mode:ESI+
Ion Spray Voltage:5 kV
Gas 1:50 psi (344.7 kPa)
Gas 2:60 psi (413.7 kPa)
Source Temp.:600 °C
InstrumentShimadzu UFLCXR