Welcome to the New Restek.com! Take a tour of the new site and feel free to send us your feedback.

Your web browser will no longer be supported by Restek.com as of 30 June 2021.
To avoid any interruption in access or functionality, install a current-generation web browser now. Learn more.

EU 15+1 PAHs on Pinnacle II PAH (FLD)

ColumnPinnacle II PAH, custom (cat.# 921946E)
Dimensions:150 mm x 3.0 mm ID
Particle Size:4 µm
Pore Size:110 Å
Temp.:30 °C
Samplecustom EU 15+1 PAH mix
Diluent:acetonitrile
Conc.:30 ng/mL (ppb)
Inj. Vol.:5 µL
Mobile Phase
A:water
B:95:5 acetonitrile:tetrahydrofuran
Time (min)Flow (mL/min)%A%B
0.001.53070
3.001.52575
10.001.50100
14.501.50100
14.511.53070
16.501.53070
Max Pressure:275 bar
DetectorShimadzu RF-10AXL
Time (min)Wavelength Ex (nm)Wavelength Em (nm)Gain
02703901x
4.853005124x
5.802904301x
8.403704604x
8.902904301x
9.603704601x
InstrumentShimadzu Prominence
NotesRF-10AXL sensitivity: medium
Cyclopenta(c,d)pyrene was included in the EU 15+1 PAH mix but it does not fluoresce.