Welcome to the New Restek.com! Take a tour of the new site and feel free to send us your feedback.
Your web browser will no longer be supported by Restek.com as of 30 June 2021.
To avoid any interruption in access or functionality, install a current-generation web browser now. Learn more.

LC Multiresidue Pesticide Standard #3 on Ultra Aqueous C18 by LC-MS/MS

LC_FF0519
PeakstR (min)Precursor Ion (m/z)Product Ion 1 (m/z)Product Ion 2 (m/z)
1.3-Hydroxycarbofuran4.5238.1163.0181.0
2.Dioxacarb4.6224.1167.0123.0
3.Fenuron4.7165.172.146.0
4.Bendiocarb5.3224.1167.1109.0
5.Propoxur5.3210.1111.0168.1
6.Carbofuran5.4222.1165.1123.0
7.Pirimicarb5.6239.272.1182.1
8.Propham5.8180.1138.0120.0
9.Fluometuron5.9233.172.146.0
10.Isoprocarb5.9194.195.0137.0
11.Monolinuron5.9215.1126.199.0
12.Forchlorfenuron6.1248.0129.193.1
13.Metobromuron6.1259.0170.2148.2
14.Chlorotoluron6.2213.172.246.2
15.Isoproturon6.4207.272.146.1
16.Fenobucarb6.5208.295.3152.1
17.Diuron6.6233.172.072.1
18.Siduron6.6233.3137.294.0
19.Diethofencarb6.7268.1226.1124.0
PeakstR (min)Precursor Ion (m/z)Product Ion 1 (m/z)Product Ion 2 (m/z)
20.Cycluron6.8199.189.189.0
21.Promecarb7.0208.1151.0109.0
22.Linuron7.1249.1182.1160.0
23.Iprovalicarb isomer 17.3321.2119.0203.1
24.Iprovalicarb isomer 27.4321.2119.0203.1
25.Bifenazate7.6301.1170.1198.1
26.Chloroxuron8.1291.172.4218.1
27.Diflubenzuron8.3311.0158.2141.1
28.Fenoxycarb8.3302.1116.188.0
29.Neburon8.4275.088.0114.0
30.Pyraclostrobin9.4388.0194.0163.0
31.Triflumuron9.4359.1156.2139.0
32.Thiobencarb9.6258.0125.089.0
33.Indoxacarb9.8528.0203.0218.0
34.Hexaflumuron10.3461.1158.2141.1
35.Novaluron10.3493.0158.1141.1
36.Teflubenzuron10.6381.1141.2158.2
37.Lufenuron10.8511.1158.1141.2
38.Flufenoxuron10.9489.0158.0141.1
39.Chlorfluazuron11.2540.0383.0158.0
Isomers are numbered sequentially based on elution order. To download this retention table in XLS format, visit www.restek.com/LC-multiresidue
ColumnUltra Aqueous C18 (cat.# 9178312)
Dimensions:100 mm x 2.1 mm ID
Particle Size:3 µm
Pore Size:100 Å
Temp.:50 °C
SampleLC multiresidue pesticide standard #3 (cat.# 31974)
Diluent:Water
Conc.:50 ng/mL
Inj. Vol.:20 µL
Mobile Phase
A:Water + 4 mM ammonium formate + 0.1% formic acid
B:Methanol + 4 mM ammonium formate + 0.1% formic acid
Time (min)Flow (mL/min)%A%B
0.000.59010
1.500.59010
4.000.54060
8.000.53070
11.000.50100
12.000.50100
12.010.59010
15.000.59010
Max Pressure:255 bar
DetectorAB SCIEX API 4000 LC-MS/MS
Ion Source:TurboIonSpray®
Ion Mode:ESI+
Ion Spray Voltage:5.5 kV
Curtain Gas:30 psi (206.8 kPa)
Gas 1:40 psi (275.8 kPa)
Gas 2:45 psi (310.3 kPa)
CAD:10 psi (68.9 kPa)
Source Temp.:350 °C
Mode:Scheduled MRM
MRM Detection Window:60 sec
Target Scan Time:0.33 sec
InstrumentAPI LC-MS/MS
NotesQ1/Q3 Resolution: Unit
Autosampler Temp.: 5 °C