Welcome to the New Restek.com! Take a tour of the new site and feel free to send us your feedback.
Your web browser will no longer be supported by Restek.com as of 30 June 2021.
To avoid any interruption in access or functionality, install a current-generation web browser now. Learn more.

Anthocyanidins on Raptor ARC-18 5 μm by HPLC

LC_FF0556
PeakstR (min)
1.Delphinidin0.724
2.Cyanidin0.992
3.Pelargonidin1.288
4.Malvidin1.442
ColumnRaptor ARC-18 (cat.# 931455E)
Dimensions:50 mm x 3.0 mm ID
Particle Size:5 µm
Temp.:40 °C
Sample
Diluent:0.1% Trifluoroacetic acid in 50:50 acetonitrile:water
Conc.:30 µg/mL
Inj. Vol.:5 µL
Mobile Phase
A:0.1% Trifluoroacetic acid in water
B:0.1% Trifluoroacetic acid in methanol
Time (min)Flow (mL/min)%B
0.001.035
2.001.055
2.011.035
4.001.035
DetectorPDA @ 530 nm
InstrumentHPLC