Welcome to the New Restek.com! Take a tour of the new site and feel free to send us your feedback.
Your web browser will no longer be supported by Restek.com as of 30 June 2021.
To avoid any interruption in access or functionality, install a current-generation web browser now. Learn more.

Food Adulterants on Raptor HILIC-Si

LC_FS0515
PeakstR (min)Precursor IonProduct IonProduct IonPolarity
1.Cyanuric acid0.47127.884.942.1-
2.Cyromazine0.52167.068.285.1+
3.Melamine0.89127.285.068.3+
4.Ammelide2.18129.186.170.2+
5.Ammeline2.97128.286.269.1+
ColumnRaptor HILIC-Si (cat.# 9310A52)
Dimensions:50 mm x 2.1 mm ID
Particle Size:2.7 µm
Pore Size:90 Å
Temp.:30 °C
Sample
Diluent:5:95 Water:acetonitrile, 10 mM ammonium formate, 0.1% formic acid
Conc.:25 ng/mL
Inj. Vol.:5 µL
Mobile Phase
A:Water, 10 mM ammonium formate, 0.1% formic acid
B:5:95 Water:acetonitrile, 10 mM ammonium formate, 0.1% formic acid
Time (min)Flow (mL/min)%A%B
0.000.60100
0.500.60100
3.500.6595
3.510.60100
8.000.60100
DetectorMS/MS
Ion Mode:ESI+/ESI-
Mode:Scheduled MRM
InstrumentHPLC