Acrylamide Reference Standard (200 ppb) and IS on Allure Acrylamide Column

LC_FS0529
PeakstR (min)Conc.
(ng/mL)
PrecursorProduct
1.Acrylamide-2,3,3-d31.73920075.158.1
2.Acrylamide1.77720072.155.1
ColumnAllure Acrylamide (cat.# 9167552)
Dimensions:50 mm x 2.1 mm ID
Particle Size:5 µm
Pore Size:60 Å
Guard Column:Allure Acrylamide 10 mm, 2.1 mm ID, 5 µm (cat.# 916750212)
Temp.:22 °C
SampleAcrylamide (cat.# 30494)
Diluent:Water
Conc.:200 ng/mL
Inj. Vol.:10 µL
Mobile Phase
A:0.001% Formic acid in water
B:0.001% Formic acid in acetonitrile
Time (min)Flow (mL/min)%A%B
0.000.41000
1.000.41000
2.000.40100
2.010.41000
5.500.41000
DetectorMS/MS
Ion Mode:ESI+
Mode:MRM