17 Polar Pesticide Panel on Raptor Polar X by LC-MS/MS

LC_FS0546
PeakstR (min)Conc.
(ng/mL)
Precursor IonProduct Ion 1Product Ion 2Product Ion 3Precursor 2Product Ion 2
1.Aminomethylphosphonic acid (AMPA)0.805200110.179.163.181.1--
2.Bialophos0.847100322.288.2216.1134.2--
3.Perchlorate2.5935101.084.95--98.983
4.Glufosinate3.376200180.285.295.1---
5.3-(Methylphophinico) propionic acid (MPPA)4.076100151.063.0107.1133.2--
6.Trifluoracetic acid (TFA)4.42320113.069.119.1---
7.2-Hydroxyethane phosphonic acid (HEPA)4.969100125.179.095.063--
8.Difluoroacetic acid (DFA)5.01820095.051.1----
9.Chlorate5.54210085.069.0--83.067.1
10.Ethephon5.564200143.1107.2---
11.Glyphosate6.113200168.163.179.1---
12.Bromide6.423200080.980.9--79.079.0
13.Bromate6.771600129.0113--127111.1
14.N-acetyl AMPA6.932200152.1110.162.9---
15.Fosetyl aluminum7.77580109.181.063.078.9--
16.Phosphonic acid8.27550081.162.9----
17.N-acetyl glufosinate9.980200222.2136.1134.159.0--
ColumnRaptor Polar X (cat.# 9311A32)
Dimensions:30 mm x 2.1 mm ID
Particle Size:2.7 µm
Pore Size:90 Å
Temp.:35 °C
Sample
Diluent:Water
Inj. Vol.:1 µL
Mobile Phase
A:Water, 0.5% formic acid
B:Acetonitrile, 0.5% formic acid
Time (min)Flow (mL/min)%A%B
0.000.53565
5.00.59010
11.50.59010
11.510.53565
130.53565
DetectorMS/MS
Ion Mode:ESI-
Mode:MRM
InstrumentUHPLC