Aromatics on Raptor Biphenyl

LC_GN0546
PeakstR (min)Conc.
(µg/mL)
1.Benzene1.63500
2.Nitrobenzene2.09100
3.1,3,5-Trinitrobenzene4.83100
4.2,4-Dinitrotoluene5.35100
ColumnRaptor Biphenyl (cat.# 9309A55)
Dimensions:50 mm x 4.6 mm ID
Particle Size:2.7 µm
Temp.:40 °C
Standard/Sample
Diluent:Acetonitrile
Conc.:100-500 µg/mL
Inj. Vol.:1 µL
Mobile PhaseWater:Methanol (50:50)
Flow:1.2 mL/min
DetectorWaters Acquity PDA @ 254 nm
InstrumentWaters ACQUITY UPLC H-Class