Mycotoxins (STD 1 - 2 ng/g) in MCT Oil on Raptor Biphenyl by LC-MS/MS

LC_GN0585
PeakstR (min)Precursor IonProduct Ion 1Product Ion 2
1.Aflatoxin G2-13C171.340348.3330.3-
2.Aflatoxin G21.344331.2189.3115.2
3.Aflatoxin G1-13C171.576346.3257.3-
4.Aflatoxin G11.579329.2243.2215.3
5.Aflatoxin B2-13C171.707332.3303.3-
6.Aflatoxin B21.711315.3287.2243.3
7.Ochratoxin A1.824404.3239.1358.3
8.Ochratoxin A-13C201.825424.3250.2-
9.Aflatoxin B11.946313.2241.2128.2
10.Aflatoxin B1-13C171.948330.3301.4-
ColumnRaptor Biphenyl (cat.# 9309A52)
Dimensions:50 mm x 2.1 mm ID
Particle Size:2.7 µm
Pore Size:90 Å
Guard Column:Raptor Biphenyl EXP guard column cartridge 5 mm, 2.1 mm ID, 2.7 µm (cat.# 9309A0252)
Temp.:35 °C
Sample
Diluent:45:55 Water:Methanol
Conc.:2 ng/g
Inj. Vol.:5 µL
Mobile Phase
A:Water, 2 mM ammonium formate, 0.1% formic acid
B:Methanol, 2 mM ammonium formate, 0.1% formic acid
Time (min)Flow (mL/min)%A%B
0.000.73565
2.000.71090
2.010.73565
3.000.73565
DetectorMS/MS
Ion Mode:ESI+
Mode:MRM
InstrumentUHPLC
NotesA working calibration solution containing Aflatoxin B1, B2, G1, and G2 and Ochratoxin A was prepared at 50.0 ng/mL in methanol. 0.25 g of MCT oil was spiked to 2 ng/g using 10 µL of the working calibration solution. A working internal standard was prepared using 13C labeled analogs at a concentration of 250 ng/mL in methanol. 10 µL of the working internal standard was aliquoted into the sample followed by vortexing for 10 seconds at 3,000 rpm. 1 mL of 45:55 H2O:MeOH was added to the sample. The sample was vortexed for 30 seconds at 3,000 rpm. The sample was then centrifuged at 3,000 xg for 5 min. at 10 °C. 750 µL of the supernatant was transferred to a conditioned (1 mL 45:55 water:methanol) Resprep bonded reversed phase SPE cartridge (Restek cat.# 26030). The sample was pulled through under vacuum into an autosampler vial for LC-MS/MS analysis.