Solvent-Savings Analysis of Cannabinoid Neutrals Mix on Raptor ARC-18 2.7 μm by LC-UV

LC_GN0672
PeakstR (min)Conc.
(µg/mL)
1.Cannabidivarin (CBDV)1.6250
2.Cannabigerol (CBG)2.2350
3.Cannabidiol (CBD)2.3650
4.Tetrahydrocannabivarin (THCV)2.5750
5.Cannabinol (CBN)3.4650
6.Delta-9-Tetrahydrocannabinol (Δ9-THC)4.3350
7.Delta-8-Tetrahydrocannabinol (Δ8-THC)4.4750
8.Cannabicyclol (CBL)5.1450
9.Cannabichromene (CBC)5.4050
ColumnRaptor ARC-18 (cat.# 9314A62)
Dimensions:150 mm x 2.1 mm ID
Particle Size:2.7 µm
Pore Size:90 Å
Guard Column:Raptor ARC-18 5 mm, 2.1 mm ID, 2.7 µm (cat.# 9314A0252)
Temp.:30 °C
SampleCannabinoid neutrals mix (cat.# 34132)
Diluent:Methanol
Conc.:50 ppm
Inj. Vol.:2 µL
Mobile Phase
A:Water, 5 mM ammonium formate, 0.1% formic acid
B:Acetonitrile, 0.1% formic acid
Time (min)Flow (mL/min)%A%B
0.000.42575
10.000.42575
DetectorUV/Vis @ 228 nm
InstrumentUHPLC