Welcome to the New Restek.com! Take a tour of the new site and feel free to send us your feedback.

Your web browser will no longer be supported by Restek.com as of 30 June 2021.
To avoid any interruption in access or functionality, install a current-generation web browser now. Learn more.

Share:

Q-sep Multispeed Centrifuge for QuEChERS

  • Program 10 custom cycles for time, braking, and speed or g-force (up to 4500 rpm or 3450 xg).
  • QuEChERS-specific presets for AOAC and EN methods make consistent operation quick and simple.
  • Convenient lid lighting indicates at a glance if unit is ready, running, or done.
  • Control panel can be temporarily locked on one cycle for error-free reproducibility.
  • Cool-Flow design prevents samples from overheating by maintaining unit at room temperature.
  • Tube holders are carbon fiber for high strength, durability, and years of trouble-free use.
  • Clear lid permits safe observation of samples and optical calibration of speed.
Documentation and Instructions Sheets

Q-sep Multispeed Centrifuge for QuEChERS

Catalog #

Select Catalog # below
Availability:
Price:
Total:
Find a Distributor
Select Catalog # below

Product Options

Not sure which Restek product is the best for your analysis? Consult a Technical Specialist
Catalog # Clear All
28295 ea. Q-sep Multispeed Centrifuge for QuEChERS Select Deselect

Catalog # 28295

Units
ea.  
Certification/Compliance
UL61010-1/CSA C22.2 No. 61010-1 and IEC61010-2-020; FDA listed; MET U.S. E112532; CE; RoHS  
Includes
15 mL four-place tube holder (6); 50 mL single-place tube holder (6); 50 mL conical tube insert (6); 2 mL tube adaptors (24); U.S. power cord (1); global/universal power cord (1)  
Product Name
Q-sep Multispeed Centrifuge for QuEChERS  
Similar To
Restek (Discontinued) 26231
Restek (Discontinued) 26230

The Q-sep centrifuge is a continuous-duty, electronically controlled, horizontal laboratory centrifuge with a lid safety interlock system. The unit is controlled by an electronic push-button timer that is variable from one to 30 minutes for precise spin times and ease of use. Samples can be safely viewed through the transparent lid. The imbalance detection system safely terminates a run cycle in the event that a load is severely imbalanced. Entry into the machine is restricted during operation by the safety interlock system. The Q-sep centrifuge features a lighted control panel that displays the status of the machine, easily viewable from a distance. The unit comes with presets for AOAC and EN methods or up to 10 custom cycles can be programmed with full control of time, braking, and speed or g-force (up to 4500 rpm or 3450 xg).

Specifications

Tube Capacity 6 x 50 mL tubes
18 x 15 mL tubes
24 x 2 mL tubes
Dimensions (H x W x D) 9 in x 14.5 in x 17 in (23 cm x 37 cm x 43 cm)
Weight 39 lb (17 kg)
Sound Level 64 dB A
Environmental Range 16–32 °C
Voltage 95–253 VAC
Frequency 50/60 Hz
Power Requirement 220 Watts
Centrifuge Motor ½ H.P. Brushless DC
Max g-Force 3450 xg
Max Speed 4500 RPM
Cycle Time 30 sec to 99 min, 59 sec (±2%)


Intended Use

General-purpose laboratory centrifuge intended for safe and rapid density-based separation of fluids, including physiologic fluids, in approved specimen receptacles for qualitative or quantitative test procedures. As a general-purpose laboratory centrifuge, it is designed to also run other approved containers filled with chemicals (nonflammable, nonexplosive, nonvolatile, and non-highly reactive only), environmental samples, and other nonhuman body samples. This device is intended to be operated by properly trained personnel who have carefully read the operating manual and are familiar with the function of the device.