Simple Large Volume Injection Method for Trace-Level Glyphosate in Bottled Drinking Water

  • Analyze glyphosate without derivatization or ion pairing.
  • Unique Raptor Polar X column provides good retention and efficient elution, resulting in better peak shape.
  • Direct large volume injection makes it easy to increase sensitivity at ppt levels.
 

The analysis of glyphosate in water is challenging primarily because it is not retained on reversed-phase columns and its response is often poor due to chelation with metal surfaces along the sample path through the LC-MS/MS system. Derivatization and ion-pairing reagents are often used to improve performance, but they can be unfriendly to the instrument and require additional sample preparation time. Using a Raptor Polar X column is a better alternative because the novel stationary phase has an affinity for polar anionic compounds, such as glyphosate, and it provides both effective retention and efficient elution.

In the method for glyphosate in water shown here, an ion-exchange mechanism in the column provides enough retention under the initial mobile phase conditions to allow large volumes (500 µL) of bottled water to be injected, which is a very simple way to improve low-level sensitivity. To achieve the best retention/elution characteristics, the mobile phase additive and pH were carefully chosen to provide a charge state that leads to adequate retention and optimal sensitivity. In addition, a one-time passivation of the LC system (without an analytical or guard column in the flow path) helped prevent undesirable chelation and a good response was observed at 100 ppt (S:N > 50).

LC_EV0571
PeakstR (min)Conc.
(ng/L)
Precursor IonProduct Ion 1Product Ion 2
1.Glyphosate6.13100168.1062.9578.95
ColumnRaptor Polar X (cat.# 9311A32)
Dimensions:30 mm x 2.1 mm ID
Particle Size:2.7 µm
Pore Size:90 Å
Temp.:30 °C
Standard/SampleGlyphosate (cat.# 32426)
Diluent:Direct injection of bottled drinking water fortified at 100 ppt with glyphosate.
Inj. Vol.:500 µL
Mobile Phase
A:Water, 0.5% formic acid
B:Acetonitrile, 0.5% formic acid
Time (min)Flow (mL/min)%A%B
0.000.5595
1.00.5595
3.00.59010
9.00.59010
9.010.5595
10.50.5595
DetectorMS/MS
Ion Mode:ESI-
InstrumentUHPLC
EVFA3196-UNV

Related Articles

Raptor Polar X: Separate a Wide Variety of Polar Analytes with a Novel Hybrid Stationary Phase

Read Article

Novel Stationary Phase for Comprehensive PFAS Analysis: Ultrashort-Chain (C2, C3), Alternative, and Legacy Compounds

Read Article

Related Products

Raptor Polar X LC Columns

View Product

LC Passivation Solution

View Product

Glyphosate Standard

View Product

Bluestem Glass Solvent Filter

View Product

Survival Kit for HPLC, PEEK

View Product