概要

特徴&利点
  • Turn in-house compressed air into ultra-pure air.
  • Remove hydrocarbons to less than 0.1 ppm by catalytic oxidation.
  • Operate at 40–125 psi (276–862 kPa).
  • Typical payback is less than one year, based on cylinder costs.
  • Installs easily and takes up little bench space.
  • Maintenance kits include a one-year supply of prefilters and final filter.
Generator installation, warranty, and service plans are also available through Restek. Contact Restek Technical Service to arrange.
製品詳細

Specifications for Parker Balston Zero Air Generators

Maximum Zero Air Flow Rate: 75-83NA
HPZA-3500
HPZA-7000
HPZA-18000
HPZA-30000
1.0 L/min
3.5 L/min
7.0 L/min
18 L/min
30 L/min
Outlet Hydrocarbon Concentration (as methane): 75-83NA
HPZA-30000
Other Models
< 0.1 ppm
< 0.1 ppm
< 0.05 ppm
Minimum/Maximum Inlet Air Pressure: 40 psig/125 psig (276/862 kPa)
Maximum Inlet Hydrocarbon Concentration (as methane): 100 ppm
Pressure Drop at Maximum Flow Rate: 4 psi (28 kPa) differential
Maximum Inlet Air Temperature: 78°F (25°C)
Inlet/Outlet Ports: 1/4" NPT (female)
Start-up Time to Specified Hydrocarbon Concentration: 45 minutes
Electrical Requirements: 75-83NA
HPZA-3500, HPZA-7000
HPZA-18000, HPZA-30000
120 VAC/60 Hz, 0.5 amps
120, 220, 230, 240 VAC/60 Hz, 2.0 amps
120, 220, 230, 240 VAC/60 Hz, 4.0 amps
Dimensions: 75-83NA
Other Models
10" h x 12" w x 3" d (25 cm x 30 cm x 8 cm)
16" h x 11" w x 13" d (42 cm x 27 cm x 34 cm)
Shipping Weight: 75-83NA
Other Models
7 lb (3 kg)
41 lb (19 kg)

Model Number of FIDs (based on a 450 cc/min fuel air rate)
75-83NA Up to 2
HPZA-3500 Up to 8
HPZA-7000 Up to 16
HPZA-18000 Up to 40
HPZA-30000 Up to 66

仕様

製品仕様
カタログ番号
20684
製品名
Parker Zero Air Generator, 75-83NA Model, 1000 cc/min Capacity, 壁掛け・ベンチトップ型