Parker Nitrogen Generator, UHPN2-1100 Model (Ultra-High Purity Zero Grade)
  1. Parker Nitrogen Generator, UHPN2-1100 Model (Ultra-High Purity Zero Grade)

Parker

Parker Nitrogen Generator, UHPN2-1100 Model (Ultra-High Purity Zero Grade)

カタログ番号 20697
ご注文時の注意点

Generator installation, warranty, and service plans are also available through Restek. Contact Restek Technical Service to arrange.

International power cords are available. Contact Customer Service to order.

Parker Nitrogen Generator, UHPN2-1100 Model (Ultra-High Purity Zero Grade)

Parker Nitrogen Generator, UHPN2-1100 Model (Ultra-High Purity Zero Grade)

概要

特徴&利点
  • Produces ultra-pure nitrogen (up to 99.9999%).
  • Requires only a compressed air source and 120 volt AC power.
  • Typical applications include GC carrier gas, make-up gas, and low-flow sample concentrators.
  • Maintenance kits include replacement filters.
  • 12-month warranty from date of purchase.
製品詳細

Specifications for Parker Balston Nitrogen Gas Generator, Model UHPN2-1100

Maximum Nitrogen Flow Rate: See Flow Table (below)
Nitrogen Purity: 99.9999%
Minimum/Maximum Inlet Pressure: 60 psig/125 psig (414/862 kPa)
Electrical Requirements1: 120 VAC/60 Hz
Dimensions: 35" h x 12" w x 16" d (89 cm x 30 cm x 41 cm)
Shipping Weight: 137 lb (62 kg)

 

Flow Table for Model UHPN2-1100

Inlet Air Pressure Maximum Outlet Flow (cc/min) Maximum Outlet Pressure
125 psig (862 kPa) 1100 85 psig (586 kPa)
110 psig (758 kPa) 1000 75 psig (517 kPa)
100 psig (689 kPa) 900 65 psig (448 kPa)
90 psig (621 kPa) 800 60 psig (414 kPa)
80 psig (552 kPa) 700 50 psig (345 kPa)
70 psig (483 kPa) 600 45 psig (310 kPa)
60 psig (414 kPa) 500 35 psig (241 kPa)

 

1Power consumption is:
Model UHPN2-1100 = 700 Watts

仕様

製品仕様
カタログ番号
20697
製品名
Parker Nitrogen Generator, UHPN2-1100 Model (Ultra-High Purity Zero Grade)