Benvenuti nel nuovo sito Restek.com! Fate un tour e sentiti liberi di inviarci un vostro riscontro

Il tuo browser web non sarà più supportato da Restek.com a partire dal 30 giugno 2021.
Per evitare qualsiasi interruzione di accesso o alle nostre funzionalità, installa subito un browser web di generazione. Per saperne di più.

Product Guide for PFAS Analysis

A Methods-Based Reference to Lab Supplies for PFAS Testing

Selecting the Right Lab Supplies for PFAS Methods

When government agencies and other research organizations developed the first PFAS analysis methods, certain per- and polyfluoroalkyl substances (PFAS) exhibited signal loss or enhancement based on how the samples were handled. These unexpected sources of error could occur anywhere in the sample pathway from collection to analysis and could lead to inaccurate quantification and reporting. Significant troubleshooting efforts from those pioneering organizations resulted in two primary observations that guide how we test these “forever chemicals” today.

  1. PFAS can be found everywhere—at the sampling site, in the analytical lab (even in instruments), and anywhere in between, so it is essential to prevent background sources from contaminating samples.
  2. PFAS can be “sticky”—they can adhere to certain materials that are commonly used in sampling and analysis (e.g., glass containers or pipettes), so everything in the sample flow path has to be considered for its ability to retain PFAS.
 

Due to these characteristics, avoiding background contamination and preventing loss due to retention are particularly important for PFAS testing. Regardless of the method you are running, best practices dictate evaluating your system and supplies prior to sample analysis to qualify them for use. In some cases, the results may warrant equipment modification or replacement of parts to ensure suitability for the analytes and LODs you are testing. The following sampling and lab supplies are often screened prior to use:

  • Sample collection vessels
  • Solid phase extraction (SPE) products
  • SPE vacuum manifolds
  • Filtration devices
  • Autosampler vials and caps
  • Mobile phase filters
  • Mobile phase transfer lines
  • PTFE tubing and filters used in UHPLC or HPLC instruments (e.g., mobile phase degasser, LC pump components, etc.)
  • Mobile phase solvents and reagents
  • LC columns
  • anything else in the sample flow path
 

Table I details the lab supplies for PFAS analysis that are used in test methods from ASTM, DIN, EPA, and ISO. When selecting the right supplies, be sure to choose method-appropriate products and evaluate your system and supplies to ensure they are free of target analytes (Table II) and qualified for PFAS analysis at the levels required by the method. Restek is proud to support PFAS testing labs with clean, high-quality lab supplies and expert technical support for the analysis of legacy, alternative, and ultrashort-chain PFAS across a wide range of global testing standards.

Table I: Lab Supplies for PFAS Analysis by Method
  U.S. EPA 537.1 U.S. EPA 533 U.S. EPA 8327 ISO 25101 ISO 21675 ASTM D7979 ASTM D7968 DIN 38407-42
Sample Matrix Drinking Water Drinking Water Various Waters & Solids (Soil, Sediment, Biota) Drinking Water, Ground Water, Surface Water Drinking Water, Natural Water, Wastewater* Water, Sludge, Influent, Effluent, Wastewater Solids (e.g., soil) Drinking Water, Ground Water, Surface Water, Wastewater, Sludge
Analytical LC Column Raptor C18 50 x 2.1 mm , 2.7 μm, (cat.#: 9304A52), SPP column
Force C18 50 x 2.1 mm, 1.8 μm (cat.# 9634252), FPP column
(Refer to our PFAS Column Selection Guide [EVAR3069B-UNV] for other column choices.)
Delay Column PFAS Delay Column (cat.#: 27854)
Sample Prep Resprep S-DVB SPE (cat.#: 28937) Resprep WAX SPE (cat.#: 28470) Resprep WAX SPE (cat.#: 28469, 28470) Resprep WAX SPE (cat.#: 28469, 28470) Resprep WAX SPE (cat.#: 28468, 28469, 28470)
Vials and Caps Polypropylene Vials (cat.#: 23242, 23245)
Polyethylene Vial Caps (cat.#: 23244, 23247)
Vacuum Manifold Resprep QR-12, QR-24 (cat.#: 28298-VM, 28299-VM)
Replacement Parts for Resprep Quick-Release (QR) SPE Vacuum Manifolds
Vacuum Pump Rocker Vacuum Pumps
Sample Reservoirs Polypropylene Sample Reservoir (cat.#: 26015)
Polypropylene Connectors (cat.#: 26007)
Polypropylene Sample Reservoir (cat.#: 26015)
Polypropylene Connectors (cat.#: 26007)
Polypropylene Sample Reservoir (cat.#: 26015)
Polypropylene Connector (cat.#: 26007)

*Containing less than 2 g/L solid particulate material.

Table II: PFAS Compound Lists by Method
U.S. EPA 537.1 U.S. EPA 533 U.S. EPA 8327
• N-ethyl perfluorooctane-sulfonamidoacetic acid (NEt-FOSAA)
• N-methyl perfluorooctane-sulfonamidoacetic acid (NMeFOSAA)
• Perfluorobutanesulfonic acid (PFBS)
• Perfluorodecanoic acid (PFDA)
• Perfluorododecanoic acid (PFDoA)
• Perfluoroheptanoic acid (PFHpA)
• Perfluorohexanesulfonic acid (PFHxS)
• Perfluorohexanoic acid (PFHxA)
• Perfluorononanoic acid (PFNA)
• Perfluorooctanesulfonic acid (PFOS)
• Perfluorooctanoic acid (PFOA)
• Perfluorotetradecanoic acid (PFTA)
• Perfluorotridecanoic acid (PFTrDA)
• Perfluoroundecanoic acid (PFUnA)
• Hexafluoropropylene oxide dimer acid (HFPO-DA)
• 9-Chlorohexadecafluoro-3-oxanone-1-sulfonic acid (9Cl-PF3ONS)
• 4,8-Dioxa-3H-perfluorononanoic acid (ADONA)
• 11-Chloroeicosafluoro-3-oxaundecane-1-sulfonic acid (11Cl-PF3OUdS)
• Perfluorobutanesulfonic acid (PFBS)
• Perfluorodecanoic acid (PFDA)
• Perfluorododecanoic acid (PFDoA)
• Perfluoroheptanoic acid (PFHpA)
• Perfluorohexanesulfonic acid (PFHxS)
• Perfluorohexanoic acid (PFHxA)
• Perfluorononanoic acid (PFNA)
• Perfluorooctanesulfonic acid (PFOS)
• Perfluorooctanoic acid (PFOA)
• Perfluoroundecanoic acid (PFUnA)
• Hexafluoropropylene oxide dimer acid (HFPO-DA)
• Perfluorobutanoic acid (PFBA)
• Perfluoropentanoic acid (PFPeA)
• Perfluoroheptanesulfonic acid (PFHpS)
• Perfluoropentanesulfonic acid (PFPeS)
• 1H,1H,2H,2H-Perfluorohexane sulfonic acid (4:2 FTS)
• 1H,1H,2H,2H-Perfluorooctane sulfonic acid (6:2 FTS)
• 1H,1H,2H,2H-Perfluorodecane sulfonic acid (8:2 FTS)
• 9-Chlorohexadecafluoro-3-oxanone-1-sulfonic acid (9Cl-PF3ONS)
• 4,8-Dioxa-3H-perfluorononanoic acid (ADONA)
• Perfluoro (2-ethoxyethane) sulfonic acid (PFEESA)
• Nonafluoro-3,6-dioxaheptanoic acid (NFDHA)
• Perfluoro-3-methoxypropanoic acid (PFMPA)
• Perfluoro-4-methoxybutanoic acid (PFMBA)
• 11-chloroeicosafluoro-3-oxaundecane-1-sulfonic acid (11Cl-PF3OUdS)
• N-ethylperfluoro-1-octanesulfonamidoacetic acid (N-EtFOSAA)
• N-methylperfluoro-1-octanesulfonamidoacetic acid (N-MeFOSAA)
• Perfluorobutanesulfonic acid (PFBS)
• Perfluorodecanoic acid (PFDA)
• Perfluorododecanoic acid (PFDoA)
• Perfluoroheptanoic acid (PFHpA)
• Perfluorohexanesulfonic acid (PFHxS)
• Perfluorohexanoic acid (PFHxA)
• Perfluorononanoic acid (PFNA)
• Perfluorooctanesulfonic acid (PFOS)
• Perfluorooctanoic acid (PFOA)
• Perfluorotetradecanoic acid (PFTeDA )
• Perfluorotridecanoic acid (PFTrA)
• Perfluoroundecanoic acid (PFUdA)
• Perfluorobutanoic acid (PFBA)
• Perfluoropentanoic acid (PFPeA)
• Perfluorodecanesulfonic acid (PFDS)
• Perfluoroheptanesulfonic acid (PFHpS)
• Perfluoropentanesulfonic acid (PFPeS)
• Perfluoro-1-octanesulfonamide (FOSA)
• 1H, 1H, 2H, 2H-perfluorohexane sulfonic acid (4:2 FTS)
• 1H, 1H, 2H, 2H-perfluorooctane sulfonic acid (6:2 FTS)
• 1H, 1H, 2H, 2H-perfluorodecane sulfonic acid (8:2 FTS)
• Perfluorononanesulfonic acid (PFNS)
 
ISO 25101 ISO 21675 ASTM D7979 & D7968 DIN 38407-42
• Perfluoro-n-octanesulfonic acid
• Perfluoro-n-octanoic acid (pentadecafluoro-noctanoic acid)
• N-ethyl perfluorooctanesulfonamidoacetic acid (N-EtFOSAA)
• N-methyl perfluorooctanesulfonamidoacetic acid (N-MeFOSAA)
• Perfluoro-n-butanesulfonic acid (PFBS)
• Perfluoro-n-decanoic acid (PFDA)
• Perfluoro-n-dodecanoic acid (PFDoDA)
• Perfluoro-n-heptanoic acid (PFHpA)
• Perfluoro-n-hexanesulfonic acid (PFHxS)
• Perfluoro-n-hexanoic acid (PFHxA)
• Perfluoro-n-nonanoic acid (PFNA)
• Perfluoro-n-octanesulfonic acid (PFOS)
• Perfluoro-n-octanoic acid (PFOA)
• Perfluoro-n-tetradecanoic acid (PFTeDA)
• Perfluoro-n-tridecanoic acid (PFTrDA)
• Perfluoro-n-undecanoic acid (PFUnDA)
• Hexafluoropropylene oxide dimer acid (HFPO-DA)
• Perfluoro-n-butanoic acid (PFBA)
• Perfluoro-n-pentanoic acid (PFPeA)
• Perfluoro-n-decanesulfonic acid (PFDS)
• Perfluoro-n-heptanesulfonic acid (PFHpS)
• Perfluorooctanesulfonamide (FOSA)
• 6:2 Fluorotelomer sulfonic acid (6:2 FTSA)
• 8:2 Fluorotelomer sulfonic acid (8:2 FTSA)
• 8:2 Fluorotelomer unsaturated carboxylic acid (8:2 FTUCA)
• 8:2 Polyfluoroalkyl phosphate diester (8:2 diPAP)
• N-methyl perfluorooctanesulfonamide (N-MeFOSA)
• N-ethyl perfluorooctanesulfonamide (N-EtFOSA)
• Perfluoro-n-hexadecanoic acid (PFHxDA)
• Perfluoro-n-octadecanoic acid (PFOcDA)
• 9-Chlorohexadecafluoro-3-oxanonane-1-sulfonic acid (9Cl-PF3ONS, aka F-53B or 6:2CL-PFESA)
• 4,8-Dioxa-3H-perfluorononanoic acid (DONA)
• Perfluorobutylsulfonate (PFBS)
• Perfluorodecanoate (PFDA)
• Perfluorodecanoate (PFDoA)
• Perfluoroheptanoate (PFHpA)
• Perfluorohexylsulfonate (PFHxS)
• Perfluorohexanoate (PFHxA)
• Perfluorononanoate (PFNA)
• Perfluorooctylsulfonate (PFOS)
• Perfluorooctanoate (PFOA)
• Perfluorotetradecanoate (PFTreA)
• Perfluorotridecanoate (PFTriA)
• Perfluoroundecanoate (PFUnA)
• Perfluorobutanoate (PFBA)
• Perfluoropentanoate (PFPeA)
• 2H-perfluoro-2-decenoic acid (FOUEA)
• Decaluoro-4-(pentafluoroethyl) cyclohexanesulfonate (PFecHS)
• 2-Perfluorohexyl ethanoic acid (FHEA)
• 2-Perfluorooctyl ethanoic acid (FOEA)
• 2-Perfluorodecyl ethanoic acid (FDEA)
• 3-Perfluoropheptyl propanoic acid (FHpPA)
• 2H-perfluoro-2-octenoic acid (FHUEA)
• Perfluoro-n-butanesulfonic acid (PFBS)
• Perfluoro-n-decanoic acid (PFDA)
• Perfluoro-n-heptanoic acid (PFHpA)
• Perfluoro-n-hexanesulfonic acid (PFHxS)
• Perfluoro-n-hexanoic acid (PFHxA)
• Perfluoro-n-nonanoic acid (PFNA)
• Perfluoro-n-octanesulfonic acid (PFOS)
• Perfluoro-n-octanoic acid (PFOA)
• Perfluoro-n-butanoic acid (PFBA)
• Perfluoro-n-pentanoic acid (PFPeA)

Access technical articles and chromatograms for PFAS analysis.

EVAR3498-UNV

Video Correlati

Analisi dei PFAS – Perché la Delay Column è importante?

Guarda il video

Articoli Correlati

PFAS LC Column Anatomy: Which Phase, Dimensions, and Particle Type Are Best?

Leggi l'articolo

Eliminate the Impact of Instrument-Related PFAS Interferences by Using a Delay Column

Leggi l'articolo

A Novel Approach for Ultrashort-Chain PFAS Analysis in Water Samples

Leggi l'articolo

Prodotti correlati

Cartucce SPE Resprep S-DVB (Fase Inversa)

Vai al prodotto

Cartucce SPE e piastre SPE da 96 pozzetti Resprep Polimeriche (Fase Inversa)

Vai al prodotto

Cartucce SPE Resprep CarboPrep Plus

Vai al prodotto

Limited-Volume 2.0 mL, 9 mm Screw-Thread Polypropylene Vials

Vai al prodotto

2.0 mL, 9 mm Solid-Top Polyethylene Caps

Vai al prodotto

Parti ed accessori per tubi SPE Resprep

Vai al prodotto

Vacuum Manifold SPE Resprep Quick-Replace (12- o 24-Porte)

Vai al prodotto

Parti di Ricambio per Vacuum Manifold SPE Resprep Quick-Replace

Vai al prodotto

Rocker Pompe da vuoto

Vai al prodotto

Delay Column PFAS

Vai al prodotto

Colonne LC Raptor C18 (USP L1)

Vai al prodotto

Filtro di aspirazione slip-on a bassa pressione per serbatoio fase mobile

Vai al prodotto

Kit di sopravvivenza per HPLC, acciaio inox

Vai al prodotto