Benvenuti nel nuovo sito Restek.com! Fate un tour e sentiti liberi di inviarci un vostro riscontro

Il tuo browser web non sarà più supportato da Restek.com a partire dal 30 giugno 2021.
Per evitare qualsiasi interruzione di accesso o alle nostre funzionalità, installa subito un browser web di generazione. Per saperne di più.

EPA Method 530 UCMR4 Standard at 10x the Method Reporting Limit on Rxi-5Sil MS (SIM)

GC_EV1465
PeakstR (min)Conc.
(µg/mL)
1.o-Toluidine-d9 (SS)5.641.0
2.o-Toluidine5.700.70
3.Quinoline-d7 (SS)8.221.0
4.Quinoline8.262.0
5.Butylated hydroxyanisole (BHA)11.503.0
6.Acenaphthene-d10 (IS)11.541.0
7.Phenanthrene-d10 (IS)15.051.0
ColumnRxi-5Sil MS, 30 m, 0.25 mm ID, 0.25 µm (cat.# 13623)
SampleMethod 530 UCMR4 standard (cat.# 572262)
Method 530 UCMR4 surrogate standard (cat.# 572265)
Method 530 UCMR4 internal standard (cat.# 572266)
Diluent:Dichloromethane
Injection
Inj. Vol.:1 µL pulsed splitless (hold 1.0 min)
Liner:Topaz 4 mm ID single taper inlet liner w/wool (cat.# 23303)
Inj. Temp.:275 °C
Pulse Pressure:30 psi (206.8kPa)
Pulse Time:1.05 min
Purge Flow:60 mL/min
Oven
Oven Temp.:70 °C (hold 1.5 min) to 200 °C at 10 °C/min to 320 °C at 7 °C/min (hold 3 min)
Carrier GasHe, constant flow
Flow Rate:1.2 mL/min
DetectorMS
Mode:SIM
SIM Program:
GroupStart Time
(min)
Ion(s) (m/z)Dwell (ms)
11.543106, 107, 112, 114 25
26.917102, 108, 129, 136 25
39.881137, 162, 164, 180 25
413.297160, 188 25
Transfer Line Temp.:280 °C
Analyzer Type:Quadrupole
Source Type:Stainless Steel
Drawout Plate:6 mm ID
Source Temp.:280 °C
Quad Temp.:180 °C
Solvent Delay Time:1.45 min
Tune Type:DFTPP
Ionization Mode:EI
InstrumentHP6890 GC & 5973 MSD
NotesThe EPA Method 530 UCMR4 standard analyte concentrations vary to simplify preparing ICAL levels based on the minimum method reporting levels.

The Method was developed on the Rtx-1701 column, but allows the use of any column with adequate selectivity. These target analytes are often included in EPA Method 8270 which uses the Rxi-5Sil MS column.