Benvenuti nel nuovo sito Restek.com! Fate un tour e sentiti liberi di inviarci un vostro riscontro

Il tuo browser web non sarà più supportato da Restek.com a partire dal 30 giugno 2021.
Per evitare qualsiasi interruzione di accesso o alle nostre funzionalità, installa subito un browser web di generazione. Per saperne di più.

LC Multiresidue Pesticide Standard #6 on Raptor ARC-18 by LC-MS/MS

LC_FF0547
PeakstR (min)Precursor IonProduct Ion 1Product Ion 2
1.Fuberidazole2.04185.0157.0156.0
2.Ethirimol2.43210.1140.098.0
3.Flutriafol3.35302.1123.170.2
4.Nuarimol4.07315.0252.081.1
5.Paclobutrazol4.13294.1125.170.2
6.Cyproconazole isomer 14.21292.2125.170.2
7.Triadimenol4.35296.170.299.1
8.Cyproconazole isomer 24.37292.2125.170.2
9.Fluquinconazole4.44376.0348.8306.9
10.Bromuconazole isomer 14.47376.0158.970.1
11.Tetraconazole4.48372.0159.070.1
12.Triticonazole4.52318.170.1124.9
13.Etaconazole isomer 14.65328.1205.0159.0
14.Epoxiconazole4.66330.0121.0101.0
15.Fenbuconazole4.66337.0125.070.1
16.Fenarimol4.67331.0268.081.0
PeakstR (min)Precursor IonProduct Ion 1Product Ion 2
17.Etaconazole isomer 24.70328.1205.0159.0
18.Flusilazole4.78316.0165.0247.0
19.Bromuconazole isomer 24.89376.0158.970.1
20.Diclobutrazol4.91328.070.059.1
21.Tebuconazole4.98308.0125.070.1
22.Penconazole5.00284.070.1159.0
23.Hexaconazole5.15314.0159.070.1
24.Metconazole5.19320.1125.070.0
25.Propiconazole5.19342.0159.069.0
26.Bitertanol5.27338.1269.270.1
27.Diniconazole5.35326.170.2159.0
28.Ipconazole isomer 15.46334.2125.170.0
29.Triflumizole5.49346.0277.960.0
30.Difenoconazole5.50406.0251.1111.1
31.Ipconazole isomer 25.55334.2125.170.0
32.Etoxazole6.16360.2177.2304.2
Isomers are numbered sequentially based on elution order.
ColumnRaptor ARC-18 (cat.# 9314A12)
Dimensions:100 mm x 2.1 mm ID
Particle Size:2.7 µm
Temp.:50 °C
SampleLC multiresidue pesticide standard #6 (cat.# 31977)
Diluent:Water
Conc.:20 ng/mL
Inj. Vol.:5 µL
Mobile Phase
A:Water + 2 mM ammonium formate + 0.2% formic acid
B:Methanol + 2 mM ammonium formate + 0.2% formic acid
Time (min)Flow (mL/min)%A%B
0.000.4955
2.000.44060
4.000.42575
6.000.40100
7.000.40100
7.010.4955
9.500.4955
DetectorWaters Xevo TQ-S
Ion Source:Waters Zspray™ ESI
Ion Mode:ESI+
Mode:MRM
InstrumentWaters
NotesAutosampler Temperature = 5 °C