ご使用のウェブブラウザは2021年6月30日をもってReste.comでサポートされなくなります。
お手数ですが、その他のウェブブラウザをインストールしてください。詳しくはこちらをご覧ください

(±) Linalool and (±) Linalyl Acetate on Rt-βDEXse

GC_FF1361
PeakstR 1tR 2
1.(±) Linalool25.68826.651
2.(±) Linalyl acetate29.16629.494
ColumnRt-βDEXse, 30 m, 0.32 mm ID, 0.25 µm (cat.# 13106)
Standard/Sample(±) Linalool
(±) Linalyl acetate
Diluent:Acetone
Conc.:1000 µg/mL
Injection
Inj. Vol.:1 µL split (split ratio 100:1)
Liner:Topaz 4.0 mm ID Precision inlet liner w/ wool (cat.# 23305)
Inj. Temp.:210 °C
Oven
Oven Temp.:40 °C (hold 1 min) to 230 °C at 2 °C/min (hold 3 min)
Carrier GasH2, constant flow
Linear Velocity:80 cm/sec @ 40 °C
DetectorFID @ 230 °C
Constant Column + Constant Make-up:51.4 mL/min
Make-up Gas Type:N2
Hydrogen flow:40 mL/min
Air flow:400 mL/min
Data Rate:50 Hz
InstrumentAgilent 7890A GC
Sample Preparation(±) Linalool and (±) linalyl acetate (neat) were dissolved in acetone to a final concentration of 1000 ppm in 2 mL, screw-thread vials (cat.# 21143), and capped with short-cap, screw-vial closures (cat.# 24495).
NotesThis is a composite chromatogram of (±) linalool and (±) linalyl acetate. (±) Linalool enantiomeric resolution was 5.9, and (±) linalyl acetate enantiomeric resolution was 2.3.