Oregon Pesticides on Rxi-5ms

GC_GN1230
PeakstR (min)Conc. (µg/mL)Mix
1.Dichlorvos4.636032591
2.Acephate5.526032591
3.Oxamyl5.976032588
4.Flonicamid6.216032588
5.Propoxur6.296032589
6.Ethoprophos6.396032591
7.Naled6.536032591
8.Dimethoate6.866032591
9.Carbofuran6.876032589
10.Diazinon7.126032591
11.Spiroxamine 17.666032587
12.Methyl parathion7.716032591
13.Carbaryl7.796032589
14.Metalaxyl7.836032587
15.Spiroxamine 27.986032587
16.Methiocarb8.026032588
17.Malathion8.106032591
18.Chlorpyrifos8.296032591
19.MGK-264 18.526032587
20.Thiamethoxam8.586032588
21.MGK-264 28.686032587
22.Fipronil8.846032588
23.Prallethrin9.026032590
24.Paclobutrazol9.286032589
25.Imidacloprid9.406032586
26.Imazalil9.656032589
27.Fludioxonil9.696032587
28.Myclobutanil9.886032589
29.Cinerin I9.886032590
30.Kresoxim-methyl9.946032587
31.Chlorfenapyr10.196032587
PeakstR (min)Conc. (µg/mL)Mix
32.Jasmolin I10.586032590
33.Pyrethrin I10.616032590
34.Pyrethrin I10.836032590
35.Trifloxystrobin11.146032587
36.Propiconazole 111.146032589
37.Propiconazole 211.286032589
38.Tebuconazole11.526032589
39.Piperonyl butoxide11.626032587
40.Spiromesifen12.006032587
41.Fenoxycarb12.166032589
42.Acetamiprid12.196032587
43.Phosmet12.206032591
44.Bifenthrin12.216032590
45.Bifenazate12.256032587
46.Etoxazole12.406032587
47.Chlorantraniliprole fragment12.476032589
48.Cinerin II12.846032590
49.Jasmolin II13.496032590
50.Pyrethrin II13.516032590
51.cis-Permethrin14.036032590
52.trans-Permethrin14.166032590
53.Pyridaben14.206032588
54.Spirotetramat14.656032586
55.Cyfluthrin14.65-14.856032590
56.Cypermethrin14.97-15.186032590
57.Boscalid15.076032587
58.Acequinocyl15.286032586
59.Etofenprox15.286032590
60.Thiacloprid15.736032588
61.Azoxystrobin16.936032587
62.Chlorantraniliprole fragment18.456032589
ColumnRxi-5ms, 30 m, 0.25 mm ID, 0.25 µm (cat.# 13423)
Standard/SampleOregon pesticides standard #1 (cat.# 32586)
Oregon pesticides standard #2 (cat.# 32587)
Oregon pesticides standard #3 (cat.# 32588)
Oregon pesticides standard #4 (cat.# 32589)
Oregon pesticides standard #5 (cat.# 32590)
Oregon pesticides standard #6 (cat.# 32591)
Diluent:Acetonitrile
Conc.:60 µg/mL
Injection
Inj. Vol.:1 µL splitless (hold 0.5 min)
Liner:Topaz 4.0 mm ID single taper inlet liner w/wool (cat.# 23447)
Inj. Temp.:250 °C
Purge Flow:5 mL/min
Oven
Oven Temp.:70 °C (hold 1 min) to 220 °C at 30 °C/min to 240 °C at 5 °C/min to 315 °C at 10 °C/min (hold 10 min)
Carrier GasHe, constant flow
Flow Rate:1.4 mL/min
DetectorThermo TSQ8000
Scan Program:
GroupStart Time
(min)
Scan Range
(amu)
Scan Rate
(scans/sec)
12.555-55010
Transfer Line Temp.:290 °C
Analyzer Type:Quadrupole
Source Temp.:330 °C
Solvent Delay Time:2.5 min
Tune Type:PFTBA
Ionization Mode:EI
Sample Preparation100 µL of each standard was combined with 400 µL acetonitrile (final volume 1 mL) for a final concentration of 60 ppm. The sample was pipetted into a 2 mL short-cap, screw-thread vial (cat.# 21143) and capped with a short-cap, screw-vial closure (cat.# 24495).
NotesSome compounds traditionally analyzed by LC were not GC amenable and/or only fragments were detected. These compounds, and the mixes that contain them, are:
Abamectin (32586)
(E)-Fenpyroximate (32586)
Spinosad (32586)
Aldicarb (32588)
Hexythiazox (32588)
Thiacloprid (32588)
Clofentezine (32589)
Daminozide (32589)