Q-sep Multispeed Centrifuge for QuEChERS

  • Program 10 custom cycles for time, braking, and speed or g-force (up to 4500 rpm or 3450 xg).
  • QuEChERS-specific presets for AOAC and EN methods make consistent operation quick and simple.
  • Convenient lid lighting indicates at a glance if unit is ready, running, or done.
  • Control panel can be temporarily locked on one cycle for error-free reproducibility.
  • Cool-Flow design prevents samples from overheating by maintaining unit at room temperature.
  • Tube holders are carbon fiber for high strength, durability, and years of trouble-free use.
  • Clear lid permits safe observation of samples and optical calibration of speed.
資料と取扱い説明書

Q-sep Multispeed Centrifuge for QuEChERS

カタログ番号

以下のカタログ番号を選択してください。
Availability:
:
:
お問い合わせ
以下のカタログ番号を選択してください。

製品選択

どのRestek製品がお客さまの分析に最適化ご不明な場合は、 テクニカルサポートへご相談ください。
この地域ではこの製品は提供されていません。 お問い合わせいただくと、 詳細をご確認いただけます。
カタログ番号 全てクリア

The Q-sep centrifuge is a continuous-duty, electronically controlled, horizontal laboratory centrifuge with a lid safety interlock system. The unit is controlled by an electronic push-button timer that is variable from one to 30 minutes for precise spin times and ease of use. Samples can be safely viewed through the transparent lid. The imbalance detection system safely terminates a run cycle in the event that a load is severely imbalanced. Entry into the machine is restricted during operation by the safety interlock system. The Q-sep centrifuge features a lighted control panel that displays the status of the machine, easily viewable from a distance. The unit comes with presets for AOAC and EN methods or up to 10 custom cycles can be programmed with full control of time, braking, and speed or g-force (up to 4500 rpm or 3450 xg).

Specifications

Tube Capacity 6 x 50 mL tubes
18 x 15 mL tubes
24 x 2 mL tubes
Dimensions (H x W x D) 9 in x 14.5 in x 17 in (23 cm x 37 cm x 43 cm)
Weight 39 lb (17 kg)
Sound Level 64 dB A
Environmental Range 16–32 °C
Voltage 95–253 VAC
Frequency 50/60 Hz
Power Requirement 220 Watts
Centrifuge Motor ½ H.P. Brushless DC
Max g-Force 3450 xg
Max Speed 4500 RPM
Cycle Time 30 sec to 99 min, 59 sec (±2%)


Intended Use

General-purpose laboratory centrifuge intended for safe and rapid density-based separation of fluids, including physiologic fluids, in approved specimen receptacles for qualitative or quantitative test procedures. As a general-purpose laboratory centrifuge, it is designed to also run other approved containers filled with chemicals (nonflammable, nonexplosive, nonvolatile, and non-highly reactive only), environmental samples, and other nonhuman body samples. This device is intended to be operated by properly trained personnel who have carefully read the operating manual and are familiar with the function of the device.