Chiral Separations 6: Essential oils on Rt-βDEXsm

6 January 2022
By
  • Jana Hepner
  • Share:

Chiral capillary GC has proven to be a convenient method for characterizing essential oils and differentiating natural flavors from those of synthetic origin. Chiral compounds from natural origins usually exist as one predominant optical isomer. Also, the inspection of enantiomeric ratios can characterize regional differences between oils. Sometimes a result of processing, the presence of racemic pairs (one-to-one ratios of each enantiomer) most often indicates adulteration or unnatural origin.

Since most chiral compounds naturally exist as one predominant isomer, resolution is more challenging, especially for components in higher concentrations. For primary constituents in essential oils, it is preferable to select a chiral column that provides a resolution factor value greater than two to overcome possible loss of resolution.

Lemon Oil

The Rt-βDEXsm is the optimum column for obtaining chiral profiles of Lemon and other citrus oils since it provides enantiomeric separation for the main terpene constituents like α- and β-pinenes, and limonene.

GC_FF1353
PeakstR 1tR 2
1.α-Pinene (±)10.54410.980
2.β-Pinene (±)12.68113.136
3.Limonene (L, D)14.82615.342
4.γ-Terpinene17.764-
5.α-Terpineol (±)33.54634.248

Figure 1: Lemon oil on Rt-βDEXsm, GC_FF1353

Rosemary Oil

Chiral constituents in this oil include α-pinene, β-pinene, camphene, limonene, linalool, camphor, terpinen- 4-ol, α terpineol, borneol, and isoborneol. These compounds are resolved optimally on the Rt- βDEXsm column.

GC_FF1355
PeakstR1tR2
1.α-Pinene (±)10.50010.941
2.Camphene (±)11.56611.952
3.β-Pinene (±)12.69013.200
4.Limonene (±)14.85015.585
5.Eucalyptol (1,8-cineole)15.837-
PeakstR1tR2
6.Linalool (±)26.01126.463
7.Camphor (±)27.49227.704
8.Terpinen-4-ol (±)30.64730.935
9.Isoborneol (±)32.25432.786
10.Borneol (±)33.00033.756
11.α-Terpineol (±)33.53934.237

Figure 2: Rosemary oil on Rt-βDEXsm, GC_FF1355

Peppermint Oil

The Rt-βDEXsm column is optimum for the separation of (+/-) α- and β-pinene, and limonene. Since menthone and menthol enantiomers are major constituents of peppermint oil, reducing the sample size to prevent overloading of these components and provide better enantiomeric resolution may be necessary. An alternative solution is to use an Rt-βDEXm or Rt-γDEXsa as a secondary column since they provide better resolution of menthol.

blog-chiral-blog-6-essential-oils-on-RT-BDEXsm-03.png

Figure 3: Peppermint oil on Rt-βDEXsm, GC_FF1356

In the next blog, we’ll look into dual column analysis of lavender oil.

Related Products

Rt-βDEXsm Columns (fused silica)

View Product

Related Articles

Chiral Separations 1: How do our GC chiral columns compare?

Read Article

Chiral Separations 2: Optimization of Chiral Separations

Read Article

Chiral Separations 3: Overloading and Tailing

Read Article

Chiral Separations 4: Separation with Rt-βDEXm, Rt-βDEXsm and Rt-βDEXse

Read Article

Chiral Separations 5: Separation with Rt-ßDEXsa, Rt-ßDEXcst and Rt-γDEXsa

Read Article