Peaks | |
---|---|
1. | 1,3-Butadiene |
2. | iso-Pentane |
3. | Methyl tert-butyl ether |
4. | Bromochloromethane |
5. | n-Hexane |
6. | Benzene |
7. | Cyclohexane |
8. | 1,4-Difluorobenzene |
9. | 2,3-Dimethylpentane |
10. | n-Heptane |
11. | Toluene |
12. | n-Octane |
13. | Chlorobenzene-d5 |
14. | Ethylbenzene |
Peaks | |
---|---|
15. | 2,3-Dimethylheptane |
16. | m-Xylene/p-Xylene |
17. | o-Xylene |
18. | n-Nonane |
19. | 4-Bromofluorobenzene |
20. | Isopropylbenzene |
21. | 1-Ethyl-3-methylbenzene |
22. | 1,3,5-Trimethylbenzene |
23. | n-Decane |
24. | 1,2,3-Trimethylbenzene/p-Isopropyltoluene |
25. | Butylcyclohexane |
26. | n-Undecane |
27. | Naphthalene |
28. | n-Dodecane |
Column | Rxi-1ms, 60 m, 0.32 mm ID, 1.00 µm (cat.# 13357) |
---|---|
Injection | splitless |
Oven | |
Oven Temp.: | 35 °C (hold 5 min) to 220 °C at 7 °C/min (hold 2 min) |
Carrier Gas | He, constant flow |
Flow Rate: | 1.2 mL/min |
Detector | MS |
---|---|
Mode: | Scan |
Transfer Line Temp.: | 240 °C |
Analyzer Type: | Quadrupole |
Source Temp.: | 230 °C |
Quad Temp.: | 150 °C |
Solvent Delay Time: | 4.0 min |
Ionization Mode: | EI |
Scan Range: | 35-250 amu |
Preconcentrator | Entech 7100 |
Instrument | HP6890 GC & 5973 MSD |
Acknowledgement | Courtesy of Mass DEP, Northeast Region |