Peaks | tR (min) | |
---|---|---|
1. | Propylene | 1.42 |
2. | Dichlorodifluoromethane (Freon® 12) | 1.46 |
3. | 1,2-Dichlorotetrafluoroethane (Freon® 114) | 1.58 |
4. | Chloromethane | 1.62 |
5. | Vinyl chloride | 1.73 |
6. | 1,3-Butadiene | 1.77 |
7. | Bromomethane | 2.02 |
8. | Chloroethane | 2.12 |
9. | Trichlorofluoromethane (Freon® 11) | 2.34 |
10. | Ethanol | 2.54 |
11. | Acrolein | 2.71 |
12. | 1,1-Dichloroethene | 2.78 |
13. | 1,1,2-Trichlorotrifluoroethane (Freon® 113) | 2.80 |
14. | Acetone | 2.82 |
15. | Carbon disulfide | 2.95 |
16. | Isopropyl alcohol | 2.99 |
17. | Acetonitrile (contaminant) | 3.05 |
18. | Methylene chloride | 3.20 |
19. | trans-1,2-Dichloroethene | 3.43 |
20. | Methyl tert-butyl ether (MTBE) | 3.44 |
21. | Hexane | 3.67 |
22. | 1,1-Dichloroethane | 3.80 |
23. | Vinyl acetate | 3.82 |
Peaks | tR (min) | |
---|---|---|
24. | 2-Butanone (MEK) | 4.29 |
25. | cis-1,2-Dichloroethene | 4.29 |
26. | Ethyl acetate | 4.34 |
27. | Bromochloromethane (IS) | 4.50 |
28. | Tetrahydrofuran | 4.52 |
29. | Chloroform | 4.60 |
30. | 1,1,1-Trichloroethane | 4.74 |
31. | Cyclohexane | 4.80 |
32. | Carbon tetrachloride | 4.88 |
33. | Benzene | 5.06 |
34. | 1,2-Dichloroethane | 5.12 |
35. | Heptane | 5.35 |
36. | 1,4-Difluorobenzene (IS) | 5.47 |
37. | Trichloroethylene | 5.69 |
38. | 1,2-Dichloropropane | 5.94 |
39. | Methyl methacrylate | 6.01 |
40. | 1,4-Dioxane | 6.02 |
41. | Bromodichloromethane | 6.20 |
42. | cis-1,3-Dichloropropene | 6.64 |
43. | 4-Methyl-2-pentanone (MIBK) | 6.80 |
44. | Toluene | 6.99 |
45. | trans-1,3-Dichloropropene | 7.26 |
46. | 1,1,2-Trichloroethane | 7.47 |
Peaks | tR (min) | |
---|---|---|
47. | Tetrachloroethene | 7.55 |
48. | 2-Hexanone (MBK) | 7.72 |
49. | Dibromochloromethane | 7.88 |
50. | 1,2-Dibromoethane | 8.01 |
51. | Chlorobenzene-d5 (IS) | 8.54 |
52. | Chlorobenzene | 8.58 |
53. | Ethylbenzene | 8.71 |
54. | m-Xylene | 8.87 |
55. | p-Xylene | 8.87 |
56. | o-Xylene | 9.32 |
57. | Styrene | 9.35 |
58. | Bromoform | 9.56 |
59. | 4-Bromofluorobenzene* | 9.99 |
60. | 1,1,2,2-Tetrachloroethane | 10.18 |
61. | 4-Ethyltoluene | 10.47 |
62. | 1,3,5-Trimethylbenzene | 10.55 |
63. | 1,2,4-Trimethylbenzene | 11.02 |
64. | 1,3-Dichlorobenzene | 11.35 |
65. | 1,4-Dichlorobenzene | 11.49 |
66. | Benzyl chloride | 11.65 |
67. | 1,2-Dichlorobenzene | 11.93 |
68. | 1,2,4-Trichlorobenzene | 13.47 |
69. | Hexachlorobutadiene | 13.59 |
70. | Naphthalene | 13.67 |
Column | Rxi-624Sil MS, 30 m, 0.32 mm ID, 1.80 µm (cat.# 13870) |
---|---|
Standard/Sample | TO-15 65 component mix (cat.# 34436) |
TO-14A internal standard/tuning mix (cat.# 34408) | |
Diluent: | Nitrogen |
Conc.: | 10.0 ppbv 200 cc injection |
Injection | Direct |
Oven | |
Oven Temp.: | 32 °C (hold 1 min) to 150 °C at 11 °C/min to 230 °C at 33 °C/min |
Carrier Gas | He, constant flow |
Flow Rate: | 2.0 mL/min |
Linear Velocity: | 51 cm/sec @ 32 °C |
Detector | MS |
---|---|
Mode: | Scan |
Transfer Line Temp.: | 230 °C |
Analyzer Type: | Quadrupole |
Source Temp.: | 230 °C |
Quad Temp.: | 150 °C |
Electron Energy: | 69.9 eV |
Solvent Delay Time: | 1.0 min |
Tune Type: | BFB |
Ionization Mode: | EI |
Scan Range: | 35 - 250 amu |
Scan Rate: | 3.32 scans/sec |
Preconcentrator | Nutech 8900DS |
Trap 1 Settings | |
Type/Sorbent : | Glass Beads |
Cooling temp: | -155 °C |
Preheat temp: | 5 °C |
Preheat time: | 0 sec |
Desorb temp: | 20 °C |
Desorb flow: | 5 mL/min |
Desorb time: | 360 sec |
Bakeout temp: | 200 °C |
Flush flow: | 120 mL/min |
Flush time: | 60 sec |
Sweep flow: | 120 mL/min |
Sweep time: | 60 sec |
Trap 2 Settings | |
Type/Sorbent: | Tenax |
Cooling temp: | -35 °C |
Desorb temp: | 190 °C |
Desorb time: | 30 sec |
Bakeout temp: | 200 °C |
Bakeout time: | 10 sec |
Cryofocuser | |
Cooling temp: | -160 °C |
Inject time: | 140 sec |
Internal Standard | |
Purge flow: | 100 mL/min |
Purge time: | 6 sec |
Vol.: | 20 mL |
ISTD flow: | 100 mL/min |
Standard | |
Size: | 200 mL |
Purge flow: | 100 mL/min |
Purge time: | 6 sec |
Sample flow: | 100 mL/min |
Instrument | HP6890 GC & 5973 MSD |
Acknowledgement | Nutech |