Peaks | tR (min) | |
---|---|---|
1. | Propylene | 1.57 |
2. | Dichlorodifluoromethane (Freon 12) | 1.60 |
3. | Chloromethane | 1.67 |
4. | 1,2-Dichlorotetrafluoroethane (Freon 114) | 1.68 |
5. | Vinyl chloride | 1.74 |
6. | 1,3-Butadiene | 1.79 |
7. | Bromomethane | 1.91 |
8. | Chloroethane | 1.98 |
9. | Ethanol | 2.16 |
10. | Trichlorofluoromethane (Freon 11) | 2.20 |
11. | Acrolein | 2.25 |
12. | Acetone | 2.29 |
13. | Acetonitrile (contaminant) | 2.29 |
14. | 1,1-Dichloroethene | 2.47 |
15. | Isopropyl alcohol | 2.49 |
16. | 1,1,2-Trichlorotrifluoroethane (Freon 113) | 2.55 |
17. | Methylene chloride | 2.63 |
18. | Carbon disulfide | 2.68 |
19. | trans-1,2-Dichloroethene | 2.97 |
20. | Methyl tert-butyl ether (MTBE) | 3.02 |
21. | 1,1-Dichloroethane | 3.13 |
22. | Vinyl acetate | 3.17 |
23. | 2-Butanone (MEK)* | 3.36 |
Peaks | tR (min) | |
---|---|---|
24. | Hexane* | 3.36 |
25. | cis-1,2-Dichloroethene | 3.50 |
26. | Ethyl acetate | 3.62 |
27. | Bromochloromethane (IS) | 3.67 |
28. | Chloroform | 3.75 |
29. | Tetrahydrofuran | 3.83 |
30. | 1,1,1-Trichloroethane | 4.09 |
31. | 1,2-Dichloroethane | 4.21 |
32. | Benzene | 4.36 |
33. | Carbon tetrachloride | 4.37 |
34. | Cyclohexane | 4.39 |
35. | 1,4-Difluorobenzene (IS) | 4.70 |
36. | Heptane | 5.00 |
37. | Trichloroethylene | 5.04 |
38. | 1,2-Dichloropropane | 5.09 |
39. | Methyl methacrylate | 5.23 |
40. | Bromodichloromethane | 5.28 |
41. | 1,4-Dioxane | 5.32 |
42. | 4-Methyl-2-pentanone (MIBK) | 5.81 |
43. | cis-1,3-Dichloropropene | 5.81 |
44. | trans-1,3-Dichloropropene | 6.33 |
45. | Toluene | 6.37 |
46. | 1,1,2-Trichloroethane | 6.55 |
Peaks | tR (min) | |
---|---|---|
47. | 2-Hexanone (MBK) | 6.77 |
48. | Dibromochloromethane | 7.09 |
49. | Tetrachloroethene | 7.19 |
50. | 1,2-Dibromoethane | 7.29 |
51. | Chlorobenzene-d5 (IS) | 7.99 |
52. | Chlorobenzene | 8.04 |
53. | Ethylbenzene | 8.32 |
54. | m-Xylene | 8.52 |
55. | p-Xylene | 8.52 |
56. | Styrene | 8.95 |
57. | o-Xylene | 8.98 |
58. | Bromoform | 9.00 |
59. | 1,1,2,2-Tetrachloroethane | 9.46 |
60. | 4-Bromofluorobenzene** | 9.67 |
61. | 4-Ethyltoluene | 10.40 |
62. | 1,3,5-Trimethylbenzene | 10.49 |
63. | 1,2,4-Trimethylbenzene | 11.00 |
64. | 1,3-Dichlorobenzene | 11.28 |
65. | Benzyl chloride | 11.39 |
66. | 1,4-Dichlorobenzene | 11.45 |
67. | 1,2-Dichlorobenzene | 11.83 |
68. | 1,2,4-Trichlorobenzene | 14.46 |
69. | Naphthalene | 14.63 |
70. | Hexachlorobutadiene | 14.89 |
Column | Rxi-5Sil MS, 30 m, 0.32 mm ID, 1.00 µm (cat.# 13654) |
---|---|
Standard/Sample | TO-15 65 component mix (cat.# 34436) |
TO-14A internal standard/tuning mix (cat.# 34408) | |
Diluent: | Nitrogen |
Conc.: | 10.0 ppbv 400 mL injection |
Injection | Direct |
Oven | |
Oven Temp.: | 32 °C (hold 1 min) to 150 °C at 9 °C/min to 230 °C at 33 °C/min |
Carrier Gas | He, constant flow |
Flow Rate: | 1.5 mL/min |
Linear Velocity: | 44 cm/sec @ 32 °C |
Detector | MS | ||||||||
---|---|---|---|---|---|---|---|---|---|
Mode: | Scan | ||||||||
Scan Program: | |||||||||
| |||||||||
Transfer Line Temp.: | 230 °C | ||||||||
Analyzer Type: | Quadrupole | ||||||||
Source Temp.: | 230 °C | ||||||||
Quad Temp.: | 150 °C | ||||||||
Electron Energy: | 69.9 eV | ||||||||
Solvent Delay Time: | 1.0 min | ||||||||
Tune Type: | BFB | ||||||||
Ionization Mode: | EI | ||||||||
Preconcentrator | Nutech 8900DS | ||||||||
Trap 1 Settings | |||||||||
Type/Sorbent : | Glass beads | ||||||||
Cooling temp: | -155 °C | ||||||||
Preheat temp: | 5 °C | ||||||||
Preheat time: | 0 sec | ||||||||
Desorb temp: | 20 °C | ||||||||
Desorb flow: | 5 mL/min | ||||||||
Desorb time: | 360 sec | ||||||||
Bakeout temp: | 200 °C | ||||||||
Flush flow: | 120 mL/min | ||||||||
Flush time: | 60 sec | ||||||||
Sweep flow: | 120 mL/min | ||||||||
Sweep time: | 60 sec | ||||||||
Trap 2 Settings | |||||||||
Type/Sorbent: | Tenax | ||||||||
Cooling temp: | -35 °C | ||||||||
Desorb temp: | 190 °C | ||||||||
Desorb time: | 30 sec | ||||||||
Bakeout temp: | 200 °C | ||||||||
Bakeout time: | 10 sec | ||||||||
Cryofocuser | |||||||||
Cooling temp: | -160 °C | ||||||||
Inject time: | 140 sec | ||||||||
Internal Standard | |||||||||
Purge flow: | 100 mL/min | ||||||||
Purge time: | 6 sec | ||||||||
Vol.: | 100 mL | ||||||||
ISTD flow: | 100 mL/min | ||||||||
Standard | |||||||||
Size: | 200 mL | ||||||||
Purge flow: | 100 mL/min | ||||||||
Purge time: | 6 sec | ||||||||
Sample flow: | 100 mL/min | ||||||||
Instrument | HP6890 GC & 5973 MSD | ||||||||
Acknowledgement | Nutech |