Peaks | |
---|---|
1. | Oxazepam |
2. | Lorazepam |
3. | Diazepam |
4. | Flunitrazepam |
5. | Prazepam |
6. | Clonazepam |
7. | Alprazolam |
Column | Rxi-5Sil MS, 30 m, 0.25 mm ID, 0.25 µm (cat.# 13623) |
---|---|
Standard/Sample | |
Diluent: | Butyl chloride |
Conc.: | 15 µg/mL |
Injection | |
Inj. Vol.: | 1 µL splitless (hold 1 min) |
Liner: | 3.5 mm taper splitless w/wool (cat.# 22286-200.1) |
Inj. Temp.: | 280 °C |
Purge Flow: | 32.2 (20:1 split) mL/min |
Oven | |
Oven Temp.: | 200 °C to 330 °C at 15 °C/min (hold 3 min) |
Carrier Gas | He, constant linear velocity |
Linear Velocity: | 50 cm/sec, 23.7 psi, 163.4 kPa @ 200 °C |
Detector | MS |
---|---|
Mode: | Scan |
Transfer Line Temp.: | 280 °C |
Analyzer Type: | Quadrupole |
Source Temp.: | 200 °C |
Electron Energy: | 70 eV |
Solvent Delay Time: | 4 min |
Tune Type: | PFTBA |
Ionization Mode: | EI |
Scan Range: | 50-350 amu |
Scan Rate: | 5 scans/sec |
Instrument | Shimadzu 2010 GC & QP2010+ MS |