Nitrosamines on Rxi-624Sil MS

GC_CF1180
PeakstR (min)Conc.
(ng/mL)
Precursor IonProduct IonCollision EnergyConfirmation Precursor IonConfirmation Product IonCollision Energy
1.N-Nitrosodimethylamine (NDMA)4.48120744218744312
2.N-Nitrosodimethylamine-d6 (NDMA-d6)4.48150804615804810
3.N-Nitrosomethylethylamine (NMEA)5.0812088421688438
4.N-Nitrosodiethylamine (NDEA)5.5612010244121025614
5.N-Nitrosoethylisopropylamine (NEIPA)5.961201164412116996
6.N-Nitroso-diisopropylamine (NDIPA)6.311281307112130886
7.Methylaniline (NMPA)6.5313110777261067714
8.N-Nitroso-di-n-propylamine (NDPA)6.621201308581301136
9.N-Nitrosomorpholine (NMOR)6.711201165610116866
10.N-Nitrosopyrrolidine (NPYR)6.79120100556100688
11.N-Nitrosopiperidine (NPIP)7.02120114696114848
12.N-Nitrosodi-n-butylamine (NDBA)7.65120158998116996
13.N-Nitroso-N-methyl-4-aminobutyric acid (NMBA)8.023651307010130985
14.Phenylbenzeneamine (NDPHA)9.36120169662216813936
ColumnRxi-624SilMS, 30 m, 0.25 mm ID, 1.4 µm (cat.# 13868)
Standard/SampleNitrosamine calibration mix, Method 521 (cat.# 31898)
N-Nitrosodimethylamine-d6 (cat.# 33910)
N-Nitrosodiphenylamine (cat.# 31429)
N-Nitroso-methylphenylamine
N-Nitroso-diisopropylamine
N-Nitrosoethylisopropylamine
N-Nitrosomorpholine
Diluent:Methanol
Injection
Inj. Vol.:1 µL split (split ratio 10:1)
Liner:Topaz 4.0 mm ID straight inlet liner w/ wool (cat.# 23444)
Inj. Temp.:280 °C
Oven
Oven Temp.:40 °C (hold 0.5 min) to 320 °C at 30 °C/min (hold 7 min)
Carrier GasHe, constant flow
Flow Rate:1.2 mL/min
DetectorMS
Mode:SIM
Transfer Line Temp.:280 °C
Analyzer Type:Quadrupole
Source Temp.:325 °C
InstrumentThermo Scientific TSQ 8000 Triple Quadrupole GC-MS
NotesFor N-nitroso-methylphenylamine (NMPA) and N-nitrosodiphenylamine (NDPHA), the nitroso groups are cleaved in the inlet to form methylaniline and phenylbenzeneamine, respectively.