Peaks | |
---|---|
1. | Tetrahydro-2-methylfuran |
2. | Butyraldehyde |
3. | Butyl ether |
4. | 1,3-Dioxolane |
5. | Ethyl acrylate |
6. | 1,4-Dioxane |
Column | Stabilwax, 60 m, 0.53 mm ID, 1.00 µm (cat.# 10658) |
---|---|
Standard/Sample | ethers |
Injection | |
Inj. Vol.: | 1.0 µL split (split ratio 7:1) |
Inj. Temp.: | 250 °C |
Oven | |
Oven Temp.: | 40 °C (hold 10 min) to 245 °C at 4 °C/min (hold 20 min) |
Carrier Gas | He, constant flow |
Linear Velocity: | 50 cm/sec @ 50 °C |
Detector | MS |
---|---|
Mode: | Scan |
Source Temp.: | 280 °C |
Ionization Mode: | EI |