Peaks | |
---|---|
1. | Methylamine |
2. | Dimethylamine |
3. | Trimethylamine |
4. | Ethylamine |
5. | Isopropylamine |
6. | tert-Butylamine |
7. | n-Propylamine |
8. | Diethylamine |
9. | sec-Butylamine |
Column | Rtx-35 Amine, 30 m, 0.53 mm ID, 1.00 µm (cat.# 11355) |
---|---|
Standard/Sample | Primary amines |
Diluent: | Water |
Conc.: | 50 ppm on-column |
Injection | |
Inj. Vol.: | 1 µL split (split ratio 10:1) |
Liner: | Splitless taper (4 mm), base deactivated Siltek (cat.# 20798-210.1) |
Inj. Temp.: | 250 °C |
Oven | |
Oven Temp.: | 35 °C (hold 5 min) |
Carrier Gas | He, constant pressure |
Linear Velocity: | 35.7 cm/sec |
Detector | FID @ 300 °C |
---|---|
Notes | Splitless taper (4 mm), base deactivated Siltek Liner obsolete, recommend catalog number 20798. |