Peaks | |
---|---|
1. | Naphthalene |
2. | Acenaphthylene |
3. | Acenaphthene |
4. | Fluorene |
5. | Phenanthrene |
6. | Anthracene |
7. | Fluoranthene |
8. | Pyrene |
Peaks | |
---|---|
9. | Benzo[a]anthracene |
10. | Chrysene |
11. | Benzo[b]fluoranthene |
12. | Benzo[k]fluoranthene |
13. | Benzo[a]pyrene |
14. | Indeno[1,2,3-cd]pyrene |
15. | Dibenzo[a,h]anthracene |
16. | Benzo[ghi]perylene |
Column | MXT-5, 30 m, 0.28 mm ID, 0.25 µm (cat.# 70224) |
---|---|
Standard/Sample | SV Calibration Mix #5 / 610 PAH Mix (cat.# 31011) |
Conc.: | 20 ng/µL |
Injection | |
Inj. Vol.: | 1.0 µL splitless (hold 1 min) |
Inj. Temp.: | 280 °C |
Oven | |
Oven Temp.: | 40 °C to 325 °C at 10 °C/min (hold 10 min) |
Carrier Gas | He, constant linear velocity |
Linear Velocity: | 50 cm/sec @ 325 °C |
Detector | MS |
---|---|
Mode: | Scan |
Source Temp.: | 325 °C |