Peaks | CAS # | |
---|---|---|
1. | Prometon | 1610-18-0 |
2. | Simazine | 122-34-9 |
3. | Atraton | 1610-17-9 |
4. | Propazine | 139-40-2 |
5. | Atrazine | 1912-24-9 |
Peaks | CAS # | |
---|---|---|
6. | Terbuthylazine | 5915-41-3 |
7. | Secbumeton | 26259-45-0 |
8. | Simetryne | 1014-70-6 |
9. | Ametryne | 834-12-8 |
10. | Prometryne | 7287-19-6 |
11. | Terbutryne | 886-50-0 |
Columns | Rtx-CLPesticides 30 m, 0.32 mm ID, 0.50 µm (cat.# 11139) |
---|---|
and Rtx-CLPesticides2 30 m, 0.32 mm ID, 0.25 µm (cat.# 11324) | |
using IP Deactivated Guard Column 5 m, 0.32 mm ID (cat.# 10044) | |
with Universal Angled “Y” Press-Tight Connector (cat.# 20403) | |
Standard/Sample | |
Conc.: | 50 pg on-column each compound |
Injection | Direct |
Liner: | DI Uniliner Liner for Agilent 5890 Packed Column GC (1.7mm ID x 3.0mm OD x 93mm) (cat.# 20964) |
Inj. Temp.: | 250 °C |
Oven | |
Oven Temp.: | 100 °C to 250 °C at 4 °C/min (hold 5 min) |
Carrier Gas | H2, constant pressure (9.65 psi, 66.5 kPa) |
Detector | NPD @ 275 °C |
---|---|
Instrument | Agilent/HP6890 GC |
Notes | HP 6890 with purged packed injection port and adaptor for an HP 5890 (cat.# 21303). |