Peaks | |
---|---|
1. | Dalapon |
2. | 3,5-Dichlorobenzoic acid |
3. | 4-Nitroanisole |
4. | Dicamba |
5. | MCPP |
6. | MCPA |
7. | Dichlorprop |
8. | 2,4-D |
9. | Pentachloroanisole |
Peaks | |
---|---|
10. | 2,4,5-TP |
11. | Chloramben |
12. | 2,4,5-T |
13. | 2,4-DB |
14. | Picloram |
15. | Bentazon |
16. | Dinoseb |
17. | DCPA methyl ester (Chlorthal-dimethyl) |
18. | Acifluorfen |
Columns | Rtx-CLPesticides 30 m, 0.32 mm ID, 0.50 µm (cat.# 11139) |
---|---|
and Rtx-CLPesticides2 30 m, 0.32 mm ID, 0.25 µm (cat.# 11324) | |
Standard/Sample | Derivatized chlorophenoxy herbicides |
Conc.: | On-column conc.: 200–20,000 pg/mL |
Injection | |
Inj. Vol.: | 1.0 µL Direct |
Liner: | Uniliner® inlet liner (cat.# 20335) |
Inj. Temp.: | 200 °C |
Oven | |
Oven Temp.: | 80 °C (hold 1 min) to 300 °C at 10 °C/min (hold 10 min) |
Carrier Gas | He, constant pressure (12.5 psi, 86.2 kPa) |
Temp.: | 80 °C |
Detector | ECD w/anode purge @ 300 °C |
---|