Your web browser will no longer be supported by Restek.com as of 30 June 2021.
To avoid any interruption in access or functionality, install a current-generation web browser now. Learn more.
Peaks | |
---|---|
1. | Nitrobenzene |
2. | 2-Nitrotoluene |
3. | 3-Nitrotoluene |
4. | 4-Nitrotoluene |
5. | Nitroglycerine |
6. | 1,3-Dinitrobenzene |
7. | 2,6-Dinitrotoluene |
8. | 2,4-Dinitrotoluene |
Peaks | |
---|---|
9. | 3,4-Dinitrotoluene (IS) |
10. | 1,3,5-Trinitrobenzene |
11. | Trinitrotoluene |
12. | PETN |
13. | RDX |
14. | 4-Amino-2,6-dinitrotoluene |
15. | 3,5-Dinitroaniline |
16. | 2-Amino-4,6-dinitrotoluene |
17. | Tetryl |
Column | Rtx-TNT2, 6 m, 0.53 mm ID, 1.50 µm (cat.# 12999) |
---|---|
Standard/Sample | 8095 Calibration Mix A (cat.# 31607) |
8095 Calibration Mix B (cat.# 31608) | |
3,4-dinitrotoluene (cat.# 31452) | |
Conc.: | 200-1000 pg per compound (on-column), estimated |
Injection | Direct |
Liner: | Siltek 1mm Uniliner (cat.# 21052-214.1) |
Inj. Temp.: | 250 °C |
Oven | |
Oven Temp.: | 80 °C (hold 1 min) to 180 °C at 10 °C/min to 300 °C at 30 °C/min (hold 3 min) |
Carrier Gas | He |
Dead Time: | 0.073 min |
Detector | ECD @ 330 °C |
---|---|
Notes | ECD with anode purge. Head pressure: 3 psi (20.7 KPa) Flow rate: 17 mL/min. @ 80°C |
We have determined your location as:
Or select your location below:
Sorry! We are unable to determine your location. Please select your location below.