Peaks | |
---|---|
1. | Dichlorvos |
2. | Hexamethylphosphoramide |
3. | Trichlorfon |
4. | mevinphos |
5. | Demeton-o |
6. | Thionazin |
7. | Ethoprop |
8. | Phorate |
9. | Naled |
10. | Tributyl phosphate (SS) |
11. | TEPP |
12. | Sulfotepp |
13. | Demeton-S |
14. | Terbufos |
15. | Fonophos |
16. | Dicrotophos |
17. | Diazinon |
Peaks | |
---|---|
18. | Disulfoton |
19. | Dioxathion |
20. | Monocrotophos |
21. | Dimethoate |
22. | Dichlorofenthion |
23. | Chlorpyrifos methyl |
24. | Ronnel |
25. | Phosphamidon isomer |
26. | Merphos |
27. | Chlorpyrifos |
28. | Fenthion |
29. | Aspon |
30. | parathion-methyl |
31. | Trichloronate |
32. | Malathion |
33. | Fenitrothion |
34. | Phosphamidon |
Peaks | |
---|---|
35. | Parathion-ethyl |
36. | Chlorfenvinphos |
37. | Prothiofos |
38. | Merphos oxone |
39. | Crotoxyphos |
40. | Stirofos |
41. | Sulprofos |
42. | Famphur |
43. | Carbophenothion |
44. | Fensulfothion |
45. | Triphenyl phosphate (SS) |
46. | Leptophos |
47. | Ethion |
48. | Phosmet |
49. | EPN |
50. | Tri-o-cresyl phosphate |
51. | Azinphos-methyl |
52. | Azinphos-ethyl |
53. | Coumaphos |
Column | Rtx-CLPesticides, 30 m, 0.32 mm ID, 0.50 µm (cat.# 11139) |
---|---|
Standard/Sample | triphenylphosphate (cat.# 32281) |
tributylphosphate (cat.# 32280) | |
8140/8141 OP Pesticide Calibration Mix A (cat.# 32277) | |
8141 OP Pesticide Calibration Mix B (cat.# 32278) | |
Conc.: | 200 ng/mL |
Injection | |
Inj. Vol.: | 1 µL splitless (hold 1 min) |
Inj. Temp.: | 250 °C |
Oven | |
Oven Temp.: | 100 °C to 180 °C at 10 °C/min (hold 2 min) to 300 °C at 18 °C/min (hold 3 min) |
Detector | FPD @ 280 °C |
---|