Column | Rtx-VMS, 60 m, 0.25 mm ID, 1.40 µm (cat.# 19916) |
---|---|
Standard/Sample | |
Diluent: | water |
Conc.: | 10 ppb in 5mL RO water (unless noted); ketones 2.5x |
Injection | purge and trap split (split ratio 20:1) |
Liner: | 1mm Split (cat.# 20973) |
Inj. Temp.: | 220 °C |
Purge and Trap | |
Instrument: | Tekmar LCS 3100 |
Trap Type: | Vocarb 3000 |
Purge: | 11 min @ ambient, flow 40 mL/min |
Dry Purge: | 1 min, flow 40 mL/min |
Desorb Preheat Temp.: | 245 °C |
Desorb: | 2 min @ 250 °C, flow 26 mL/min |
Bake: | 8 min @ 260 °C |
Interface Connection: | injection port |
Transfer Line Tubing: | Silcosteel Treated 0.53mm ID tubing(cat.# 20595) |
Transfer Line Temp.: | 120 °C |
Oven | |
Oven Temp.: | 60 °C (hold 2 min) to 180 °C at 12 °C/min to 225 °C at 45 °C/min (hold 6 min) |
Carrier Gas | He, constant flow |
Flow Rate: | 1.3 mL/min |
Detector | MS |
---|---|
Mode: | Scan |
Transfer Line Temp.: | 280 °C |
Analyzer Type: | Quadrupole |
Tune Type: | BFB |
Ionization Mode: | EI |
Scan Range: | 35-300 amu |
Instrument | HP6890 GC & 5973 MSD |
Notes | For proper flows, adjust retention time of dichlorodifluoromethane to a retention time of 4.03 minutes @ 60C. |