Peaks | tR (min) | |
---|---|---|
1. | N-Nitrosodimethylamine | 3.38 |
2. | Pyridine | 3.37 |
3. | Methyl methanesulfonate | 4.53 |
4. | 2-Fluorophenol (SS) | 4.61 |
5. | Ethyl methanesulfonate | 5.16 |
6. | Phenol | 5.51 |
7. | Phenol-d6- (SS) | 5.50 |
8. | Aniline | 5.52 |
9. | bis(2-chloroethyl)ether | 5.58 |
10. | 2-chlorophenol-d4 (SS) | 5.60 |
11. | 2-Chlorophenol | 5.61 |
12. | 1,3-Dichlorobenzene | 5.73 |
13. | 1,4-Dichlorobenzene-D4 (IS) | 5.79 |
14. | 1,4-Dichlorobenzene | 5.80 |
15. | Benzyl Alcohol | 5.96 |
16. | 1,2-Dichlorobenzene-D4 (SS) | 5.94 |
17. | 1,2-Dichlorobenzene | 5.96 |
18. | 2-Methylphenol | 6.06 |
19. | Bis(2-Chloroisopropyl)ether | 6.06 |
20. | 4-Methylphenol/3-Methylphenol | 6.22 |
21. | N-Nitrosodi-N-propylamine | 6.24 |
22. | Acetophenone | 6.20 |
23. | Hexachloroethane | 6.25 |
24. | Nitrobenzene-D5 (SS) | 6.35 |
25. | Nitrobenzene | 6.37 |
26. | Isophorone | 6.59 |
27. | 2,4-Dimethylphenol | 6.70 |
28. | 2-Nitrophenol | 6.66 |
29. | Benzoic acid | 6.87 |
30. | Bis(2-chloroethoxy)methane | 6.79 |
31. | 2,4-Dichlorophenol | 6.89 |
32. | 1,2,4-Trichlorobenzene | 6.95 |
33. | Naphthalene-D8 (IS) | 7.00 |
34. | Naphthalene | 7.03 |
35. | 2,6-Dichlorophenol | 7.10 |
Peaks | tR (min) | |
---|---|---|
36. | 4-Chloroaniline | 7.12 |
37. | Hexachloropropene | 7.10 |
38. | Hexachlorobutadiene | 7.14 |
39. | 4-Chloro-3-methylphenol | 7.57 |
40. | Isosafrole | 7.61 |
41. | 2-Methylnaphthalene | 7.69 |
42. | 1-Methylnaphthalene | 7.79 |
43. | Hexachlorocyclopentadiene | 7.85 |
44. | 1,2,4,5-Tetrachlorobenzene | 7.86 |
45. | 2,4,6-Trichlorophenol | 7.98 |
46. | 2,4,5-Trichlorophenol | 8.02 |
47. | 2-Fluorobiphenyl (SS) | 8.05 |
48. | Safrole | 8.11 |
49. | 2-Chloronaphthalene | 8.17 |
50. | 2-Nitroaniline | 8.32 |
51. | 1,4-Naphthoquinone | 8.37 |
52. | Dimethyl phthalate | 8.48 |
53. | 1,3-Dinitrobenzene | 8.54 |
54. | 2,6-Dinitrotoluene | 8.56 |
55. | Acenaphthylene | 8.57 |
56. | 3-Nitroaniline | 8.68 |
57. | Acenaphthene-d10 (IS) | 8.71 |
58. | Acenaphthene | 8.74 |
59. | 2,4-Dinitrophenol | 8.82 |
60. | 4-Nitrophenol | 8.91 |
61. | Pentachlorobenzene | 8.89 |
62. | 2,4-Dinitrotoluene | 8.94 |
63. | Dibenzofuran | 8.91 |
64. | 2,3,4,6-Tetrachlorophenol | 9.05 |
65. | Diethyl Phthalate | 9.15 |
66. | 4-Chlorophenyl phenyl ether | 9.24 |
67. | Fluorene | 9.24 |
68. | 4-Nitroaniline | 9.30 |
69. | 4,6-Dinitro-2-methylphenol | 9.33 |
70. | Diphenylamine | 9.37 |
Peaks | tR (min) | |
---|---|---|
71. | Azobenzene | 9.39 |
72. | 2,4,6-Tribromophenol (SS) | 9.49 |
73. | Phenacetin | 9.70 |
74. | 4-Bromophenyl phenyl ether | 9.71 |
75. | Hexachlorobenzene | 9.80 |
76. | Pentachlorophenol | 9.99 |
77. | Pentachloronitrobenzene | 10.02 |
78. | Dinoseb | 10.16 |
79. | Phenanthrene-D10 (IS) | 10.16 |
80. | Phenanthrene | 10.18 |
81. | Anthracene | 10.23 |
82. | di-n-Butyl phthalate | 10.75 |
83. | 4-nitroquinoline-1-oxide | 11.04 |
84. | Isodrin | 11.29 |
85. | Fluoranthene | 11.50 |
86. | Benzidine | 11.71 |
87. | Pyrene | 11.80 |
88. | Aramite | 12.10 |
89. | p-Terphenyl-d14 (SS) | 12.02 |
90. | Chlorbenzilate | 12.31 |
91. | Butyl benzyl phthalate | 12.84 |
92. | Kepone | 12.81 |
93. | Bis(2-ethylhexyl)phthalate | 13.99 |
94. | Chrysene-D12 (IS) | 13.87 |
95. | 3,3'-Dichlorobenzidine | 13.87 |
96. | Benzo[a]anthracene (IS) | 13.84 |
97. | Chrysene | 13.93 |
98. | Di-n-octyl phthalate | 15.51 |
99. | Benzo[b]fluoranthene | 16.29 |
100. | Benzo[k]fluoranthene | 16.35 |
101. | Benzo[a]pyrene | 17.07 |
102. | Perylene-D12 (IS) | 17.22 |
103. | 3-Methylcholanthrene | 17.98 |
104. | Indeno[1,2,3-cd]pyrene | 19.82 |
105. | Dibenzo[a,h]anthracene | 19.91 |
106. | Benzo[ghi]perylene | 20.44 |
Column | Rtx-5Sil MS, 30 m, 0.25 mm ID, 0.25 µm (cat.# 12723) |
---|---|
Standard/Sample | Semivolatiles 8270 Standard |
Conc.: | 80 ng |
Injection | splitless |
Inj. Temp.: | 300 °C |
Oven | |
Oven Temp.: | 40 °C (hold 2 min) to 245 °C at 25 °C/min to 330 °C at 6 °C/min (hold 5 min) |
Carrier Gas | He, constant flow |
Flow Rate: | 1.0 mL/min |
Detector | MS |
---|---|
Mode: | Scan |
Source Temp.: | 280 °C |
Scan Range: | 35-550 amu |