Organochlorine Pesticides US EPA Method 8081 on Rtx-CLPesticides (0.53mm ID)

ColumnRtx-CLPesticides, 30 m, 0.53 mm ID, 0.50 µm (cat.# 11140)
Standard/SampleOrganochlorine Pesticide Mix AB #2 (cat.# 32292)
Conc.: 16-160pg on-column concentration
InjectionDirect
Liner:Uniliner liner (cat.# 20335)
Inj. Temp.:200 °C
Oven
Oven Temp.:120 °C (hold 1 min) to 300 °C at 9 °C/min (hold 10 min)
Carrier GasHe, constant flow
Flow Rate:2.83 mL/min @ 120 °C
Dead Time:1.9 min
DetectorECD @ 300 °C
NotesECD with anode purge
Head pressure: 3 psi (constant)

See also: Organochlorine Pesticides US EPA Method 8081 on Rtx-CLPesticides2 (0.53mm ID) GC_EV00547B