Peaks | CAS # | |
---|---|---|
1. | Trifluralin | 1582-09-8 |
2. | Desisopropylatrazine | 1007-28-9 |
3. | Desethyl-atrazine | 6190-65-4 |
4. | Desethyl-terbuthylazine | -------- |
5. | Simazine | 122-34-9 |
6. | Atrazine | 1912-24-9 |
7. | Terbumeton | 33693-04-8 |
8. | Terbuthylazine | 5915-41-3 |
9. | Secbumeton | 26259-45-0 |
10. | Cyanazine | 21725-46-2 |
Column | Rtx-OPPesticides2, 30 m, 0.25 mm ID, 0.25 µm (cat.# 11243) |
---|---|
Standard/Sample | custom standard |
Conc.: | 100 ng/µL each component |
Injection | |
Inj. Vol.: | 1.0 µL splitless (hold 0.5 min) |
Liner: | 4mm Splitless (cat.# 20904) |
Inj. Temp.: | 300 °C |
Oven | |
Oven Temp.: | 80 °C (hold 1 min) to 140 °C at 25 °C/min (hold 5 min) to 165 °C at 1 °C/min to 300 °C at 15 °C/min (hold 1 min) |
Carrier Gas | He, constant pressure |
Linear Velocity: | 27.22 cm/sec @ 80 °C |
Detector | Agilent 5971A GC/MS |
---|---|
Transfer Line Temp.: | 280 °C |
Analyzer Type: | Quadrupole |
Ionization Mode: | EI |
Scan Range: | 35-360 amu |