Peaks | |
---|---|
1. | 2,4,5,6-Tetrachloro-m-xylene (surr.) |
2. | α-BHC |
3. | γ-BHC |
4. | β-BHC |
5. | δ-BHC |
6. | Heptachlor |
7. | Aldrin |
8. | Heptachlor epoxide |
9. | γ-Chlordane |
10. | α-Chlordane |
11. | Endosulfan I |
Peaks | |
---|---|
12. | 4,4'-DDE |
13. | Dieldrin |
14. | Endrin |
15. | 4,4'-DDD |
16. | Endosulfan II |
17. | Endrin aldehyde |
18. | 4,4'-DDT |
19. | Endosulfan sulfate |
20. | Methoxychlor |
21. | Endrin ketone |
22. | Decachlorobiphenyl (surr.) |
Column | Rtx-440, 30 m, 0.32 mm ID, 0.50 µm (cat.# 12939) |
---|---|
using Siltek Guard Column 5 m, 0.32 mm ID (cat.# 10027) | |
with SeCure “Y” Connector Kit (cat.# 20277) | |
Standard/Sample | Organochlorine Pesticide Mix AB #2 (cat.# 32292) |
Pesticide Surrogate Mix, EPA 8080, 8081 (cat.# 32000) | |
Conc.: | Organochlorine Pesticide Mix AB #2, 8-80µg/mL each component in hexane/toluene (cat.# 32292), Pesticide Surrogate Mix, 200µg/mL each component in acetone (cat.# 32000) |
Injection | |
Inj. Vol.: | 0.5 µL splitless (hold 0.75 min) |
Liner: | 3.5 mm Single Gooseneck (cat.# 20961) |
Inj. Temp.: | 250 °C |
Oven | |
Oven Temp.: | 110 °C (hold 0.5 min) to 230 °C at 25 °C/min to 330 °C at 15 °C/min (hold 1 min) |
Carrier Gas | H2, constant pressure |
Linear Velocity: | 60 cm/sec @ 300 °C |
Detector | ECD @ 330 °C |
---|---|
Instrument | Shimadzu 2010 GC |